Chapter 3 Product Specifications
41
3.15 CO
2
Specifications
Measurement technique
Infrared absorption technique
Measurement mode
Sidestream, microstream or mainstream (optional)
Displayed parameter
EtCO2, FiCO2, Respiration Rate
CO2 function
Meet the requirements of EN864 and ISO9918.
1.1.4. Mindray CO
2
Specifications
CO2 measurement range
0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2mmHg
±5%
±10%
Deflation rate
100, 150ml/min
Precision of deflation rate
15%
Start-up time of CO2
module
< 1min.
The module enters the warming up status after the startup. Ten
minutes later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time
< 240 ms (10 to 90%)
Delay time
< 2s (Length of sampling tube: 7 feet; inner diameter: 0.055
inches; sampling flow: 150ml/min)
Apnea alarm delay
AwRR: 10 to 40 s
Содержание PM-7000
Страница 1: ...PM 7000 Patient Monitor Service Manual ...
Страница 2: ......
Страница 58: ...Chapter 3 Product Specifications 46 line 2 5m ENF 350 ms Delay time 4s ...
Страница 81: ......
Страница 82: ...P N 7001 20 67407 1 0 ...