5-2
5.1.2 Data Adding
The previously stored ID number appears on the screen when a finger is inserted into the
SpO
2
sensor. It will keep blinking for 8 seconds after the pulse is found.
Press the Confirm ID button before the ID number stops blinking, DPM1 will set the
ID number as current patient ID. The data measured thereafter will be superadded to
the previous ID.
If the user doesn’t press the Confirm ID button before the ID number stops blinking, a
new ID number, which is the blinking number plus 1, will be set as the current patient
ID.
The Confirm ID button is of no use when the current ID number is 000, which will
automatically change to 001 when the pulse is found.
5.1.3 Data Protection
DPM1 has data protection function. When the power is turned off accidentally during the
process of storing a data, DPM1 will evaluate completeness of the last stored data when it is
restarted. If the data is complete, it will be validated, otherwise it will be invalidated.
5.1.4 Data Deletion
Press the Delete ID button in the standby mode, the message “DELETE ALL?” will be
displayed as shown in Figure 5-1.
Figure 5-1 Figure 5-2
To delete all stored data:
Press the Delete ID button again. As shown in Figure 5-2, the message “ALL DELETED”
will be displayed for 2 seconds. Then DPM1 will switch back to the previous standby mode,
the ID number restores to “000”, symbols of memory full as well as ID Full disappears.
Not to delete all stored data:
Don’t press the Delete ID button and wait for 10 seconds, the “DELETE ALL?” message
will disappear automatically and the previous operation will be cancelled. DPM1 will
switch back to the previous mode.