1.1.1 Warnings
WARNING
z
DPM1 is intended for use by qualified clinical physicians or well-trained nurses in
hospital.
z
DPM1 is designed for real-time and rapid measurement of the patient’s SpO
2
. It is
not suitable for long-time continuous patient monitoring. Continual measurement
must not exceed 2 hours.
z
DPM1 is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms. It is not intended as a device used
for treatment purposes.
z
To ensure patient safety, verify this device and accessories can function safely and
normally before use.
z
EXPLOSION HAZARD: Do not use DPM1 in the presence of flammable
anesthetics, explosive substances, vapors or liquids.
z
Failure to implement a satisfactory maintenance schedule may cause undue device
failure and possible health hazard.
z
When using DPM1 together with the electrical surgery equipment, the user should
pay attention to and guarantee safety of the patient being measured.
z
Do not pull or lift DPM1 by its connection cable. That may lead to falling and
consequent patient injuries.
z
Magnetic and electrical fields are capable of interfering with the proper
performance of DPM1. For this reason make sure that all external devices
operated in the vicinity of DPM1 comply with the relevant EMC requirements.
Mobile phone, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
z
DPM1 and its accessories may be contaminated by microorganism during
transporting, use and storage. Use the recommended methods to sterilize and
disinfect DPM1 or its accessories when the packing material is damaged, or it has
not been used for a long time.