5-1
5
Other Functions
5.1 Data Management
5.1.1 Data Storage
DPM1 oximeter has its internal memory to store data. The memory is divided into the ID
Data Zone and Trend Data Zone.
The ID Data Zone is capable of storing 100 patients’ ID data at most. When the number
exceeds 100, new data will automatically cover the old one from the earliest stored data.
ID001 data ID002 data ID003 data
……
ID100 data
One ID data include:
Number of the SpO2 and pulse rate values stored to this ID.
Maximum SPO2 value of this ID.
Minimum SPO2 value of this ID.
Average SPO2 value of this ID.
Maximum PR value of this ID.
Minimum PR value of this ID.
Average PR value of this ID.
Trend Data Zone is capable of storing 200 trend data (TD, as shown below) at most. When
the number exceeds 200, new data will automatically cover the old data from the earliest
stored data.
TD 001
TD 002
TD 003
……
TD 200
The first trend data will be stored 15 seconds after the pulse is found. Thereafter one trend
data will be stored every 2 minutes. One trend data includes:
Average SpO2 value within the 2 minutes.
Average PR value within the 2 minutes.
NOTE
z
The first trend data are the instant values of SpO2 and PR.
z
For a patient, his (her) ID data are calculated from all measured trend data of him
(her).
z
Once the trend data of a patient is covered, all trend data of that patient will be
deleted, however, the ID data will not be deleted.