Accutorr V Operating Instructions
3 - 39
Operation
Recorder
3.13
Recorder
The Accutorr V provides a permanent record of patient data using the recorder.
• To print the data on the current display (5 lines of data), press
(33) for less than 2
seconds (1 beep tone).
• To print all the trend data for the current patient (up to 11 lines of data per block.), press
(33) for more than 2 seconds (1 beep tone) in trend data review mode.
• To stop printing while in progress, press
(33) (1 beep tone).
FIGURE 3-24
Example Recorder Strip of Data on the Current Display
• Whenever SpO
2
is active, it is used to measure Pulse Rate.
• If data is invalid or not available for any given parameter, “—” prints under that
parameter.
• If an NIBP measurement was interrupted, “XXX” is printed under Sys, Dia, and Map.
• Parenthesis ( ) indicate a parameter value that violated alarm limits.
Time
Sys (mmHg) Dia (mmHg) Map (mmHg) SpO2 (%) PR (BPM) Temp (°F)
12 – 10 11:19
XXX
XXX
XXX
97.2
Trend Table 12 – 10 11:19
—
—
—
94
80
96.8
(ALL)
12 – 10 11:18
—
—
—
(85)
(122)
95.7
12 – 10 11:18
—
—
—
92
110
93.4
12 – 10 11:17
120
80
93
—
72
83.5
PATIENT ID: 20081210095209 Record Time: 12-10-2008 11:19:29 Mindray DS USA, Inc.
Parenthesis print around measurements that
caused an alarm violation.
When no information is available for a
particular parameter, a dash prints.
An interrupted NIBP measurement prints as
XXX for Sys, Dia, and Map.
If the temperature is shown in every time slot, the
temperature parameter is in continuous mode.
Содержание Accutorr V
Страница 1: ...Operating Instructions 0070 01 0699 02_Acctr V ops color indd 1 2 18 11 9 08 AM...
Страница 2: ...Operating Instructions 0070 02 0699 02_Acctr V ops b_w indd 1 2 18 11 9 21 AM...
Страница 98: ...Accutorr V Operating Instructions 0070 10 0699 02 5 1 5 0 Accutorr V Accessories Accessories 5 2...
Страница 129: ...0070 10 0699 02 Rev 12 0 March 2015...