Front Panel
Controls and Indicators
2 - 6
Accutorr V Operating Instructions
26. INTERVAL key
Changes the NIBP measuring mode and interval by cycling through the modes and intervals
displayed in the NIBP Interval indicator (27), as follows:
OFF (manual), STAT, or 1, 2, 3, 5, 10, 15, 20, 30, 60, 120, 240 minutes
Pressing and holding the Interval key for 3 seconds directly goes to OFF, i.e. the manual
mode.
27. NIBP Interval indicator
Indicates the current NIBP measuring mode or interval.
28. Pulse strength indicator
Indicates the patient’s relative pulse strength by the number of stacked bars.
29. SILENCE key
A quick press of this key pauses the current alarm for two (2) minutes, after which alarm tone
resumes if alarm limits are still violated. If a new alarm condition occurs during the two (2)
minutes, a new alarm tone sounds. Pressing and holding this key for more than two (2)
seconds disables alarm tones indefinitely. If a new alarm condition occurs while in this state,
the monitor automatically exits the alarm silenced state.
30. Silence indicator
•
Dark
(Normal state): when an alarm occurs, the monitor presents an alarm tone,
visual indication, and message according to the alarm level.
•
Illuminated
: Alarm silenced state: when an alarm occurs, the monitor presents a
visible alarm and alarm message, but no alarm tone is given. If a new alarm condition
occurs, the monitor automatically exits the alarm silenced state.
•
Flash
(Alarm paused status): when an alarm occurs, the monitor displays a visible
alarm and alarm message, but no alarm tone is given. The alarm paused time is 120
seconds, after which the alarm tone sounds again if alarm limits are still violated. The
unit counts down the 120 seconds on the LCD Display (17) in place of the date and
time. If a new alarm occurs during this period, the monitor automatically exits the
alarm paused state.
31. Temperature site
The temperature measuring position and monitoring mode, oral, auxiliary, and rectal
selection illuminates.
32. Temperature Unit indicator
The current temperature unit.
33. PRINT key
Starts or stops the recorder.
34. NIBP connector
Used to attach the specified NIBP hose to the Accutorr V.
Содержание Accutorr V
Страница 1: ...Operating Instructions 0070 01 0699 02_Acctr V ops color indd 1 2 18 11 9 08 AM...
Страница 2: ...Operating Instructions 0070 02 0699 02_Acctr V ops b_w indd 1 2 18 11 9 21 AM...
Страница 98: ...Accutorr V Operating Instructions 0070 10 0699 02 5 1 5 0 Accutorr V Accessories Accessories 5 2...
Страница 129: ...0070 10 0699 02 Rev 12 0 March 2015...