Patient Setup
Operation
3 - 8
Accutorr V Operating Instructions
To select the Patient Size, press
(23). Three choices are available: Adult, Pediatric,
and Neonate. The patient size changes with each key press.
The Patient size indicator (21) illuminates to indicate the selected size as shown in
FIGURE 3-6. The factory default Patient Size setting is Adult.
NOTE:
Selecting Neonate patient size changes Temperature site
(31) to Axiliary. The temperature site will default to oral
when Adult patient size is selected.
3.5.3
Setting Initial Cuff Inflation Pressure
The initial cuff inflation pressure depends on the Patient Size setting. The initial cuff inflation
pressures are listed in the following table.
NOTE:
The default patient size and initial cuff inflation pressure can
be customized.
To modify the initial cuff inflation pressure:
1.
Press
(12) to display the SYSTEM SETUP dialog as shown in FIGURE 3-7.
2.
Press
MAINTENANCE
.
FIGURE 3-7
SYSTEM SETUP Dialog
3.
Once
MAINTENANCE
is highlighted, press
(18) to display the MAINTENANCE
dialog as shown in FIGURE 3-8.
PATIENT SIZE
SETTING
FACTORY DEFAULT
INITIAL CUFF
INFLATION VALUES
LOWEST
SELECTABLE
PRESSURE
HIGHEST
SELECTABLE
PRESSURE
INCREMENT
Adult
180 mmHg
100 mmHg
280 mmHg
5 mmHg
Pediatric
140 mmHg
60 mmHg
180 mmHg
5 mmHg
Neonate
100 mmHg
40 mmHg
120 mmHg
5 mmHg
Содержание Accutorr V
Страница 1: ...Operating Instructions 0070 01 0699 02_Acctr V ops color indd 1 2 18 11 9 08 AM...
Страница 2: ...Operating Instructions 0070 02 0699 02_Acctr V ops b_w indd 1 2 18 11 9 21 AM...
Страница 98: ...Accutorr V Operating Instructions 0070 10 0699 02 5 1 5 0 Accutorr V Accessories Accessories 5 2...
Страница 129: ...0070 10 0699 02 Rev 12 0 March 2015...