36
MW150 & MW151 Bench Meters
Export Data
PC
1. Connect the
MW151
to the PC using the supplied USB cable.
2. Turn the meter on with the ON/OFF button.
3. The PC will detect the bench meter as a removable flash drive.
4. If you connect the meter to the PC, the default setting is EXPORT TO PC mode.
5. Use a file manager (e.g. Windows Explorer, MacOS Finder) to move the files from the
meter to the PC.
6. When the USB cable is attached to the PC and EXPORT TO PC mode is enabled, logging
is not possible.
Note:
To enable logging while connected to the PC, enter SETUP and change “EXPORT
TO PC
”
mode by pressing CAL/EDIT and using the arrow keys to “LOG ON METER
”
mode.
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
The .csv file (comma separated values) may be opened with text editor or spreadsheet
application.
Note:
Field separator may be set as comma or semicolon, depending on region
preferences. Western Europe (ISO-8859-1) character set and English language are
suggested settings. Other files may be visible depending upon computer settings.
Files with interval lots are named as PHLOT, followed by lot number, e.g. PHLOT001,
PHLOT002. Manual logs are named as PHLOTMAN, and stability logs are named as
PHLOTSTAB.
Notes:
If “°C!” appears in logged data, the electrode/probe was used beyond its operation
specifications and the data is not considered reliable.
If “°C!!” appears in logged data, the meter was in MTC mode.