14
MW150 & MW151 Bench Meters
6. SETUP
Setup mode allows parameters that are not directly related to the measurement to be
viewed and modified.
1. Press SETUP to enter in Setup mode.
2. Use
keys to select required parameter.
3. Press the CAL/EDIT to enter Edit mode.
4. If various options are available, use RANGE/
to select desired option.
5. Use
keys to select or change parameter values.
6. Press GLP/ACCEPT to confirm or press CAL/EDIT to exit without changing.
7. Once in Edit mode, below detailed parameters can be set or viewed.
If needed, use ESC to exit Setup mode.
6.1. GENERAL SETUP
pH Information
Use
keys to select whether the pH buffer calibration information is enabled or not:
on – enable (default), off – disable.
When enabled, the electrode segment displays the electrode condition (
MW151
only).
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
Calibration Expired Warning
Use
keys to select the number of days since last calibration has elapsed: 1 to 7 days
(default), or off.
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME