Instruction Manual Mi451 COPPER Photometer for wine analysis
1 2
1 2
1 2
1 2
1 2
• Remove the cuvet from the instrument.
• Insert the reacted sample cuvet into the holder
and close the lid.
• Press
READ
and the display will show “----”
during measurement.
• The instrument directly displays concentration
in mg/L (ppm) of copper on the Liquid Crystal
Display.
Note
If the copper concentration exceeds 1.50 ppm
or if the sample is very turbid or dark red
colored, it is recommended to dilute the
sample 10 times with
Mi550S1-0
Wine
Solvent-1 and repeat the complete
measurement procedure starting from the
beginning, taking 15 mL of diluted wine sample
in vial A for zero-sampling and 15 mL in vial
B for reading. In this case the displayed value
needs to be multiplied by 10 to compensate
for dilution.