Indicates that the batteries are dead and must be
replaced. After this message appears, the instrument
is switched off. Change the batteries and restart the
meter.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure
best accuracy.
• Use the supplied plastic pipette for adding the exact amount
of wine sample (15 mL) or reagents A or B (up to the 20 mL
mark) to the graduated glass vials. The liquid in the vial
forms a convexity on the top; the bottom of this convexity
must be at the same level of the mark.
• To transfer the supernatant organic solvent from the vials to
the cuvets use two different clean 1 mL pipettes, one for the zero and one for the
sample.
a) squeeze the bulb of the pipette;
b) insert the plastic pipette into the supernatant organic solvent and release the
bulb slowly, paying attention not to transfer the wine sample too;
c) fill a cuvet with organic solvent by squeezing the bulb of the pipette.