background image

www.milwaukeetesters.com

9

• Diluting procedure of the wine sample in case the “L Lo” (Low

Light) message appears: use the pipette to fill the glass vials with
5 mL of wine sample, then fill the vials up to the 15 mL mark with

Mi550S1-0

. This is the diluted wine sample. The final reading

must be multiplied by 3 to compensate for dilution.

• In order to avoid reagent leaking and to obtain more accurate

measurements, it is recommended to close the cuvet first with the

supplied HDPE plastic stopper 

and then with the black cap.

• Whenever the cuvet is placed into the measurement cell, it must

be dry outside, and completely free of fingerprints, oil or dirt.
Wipe it thoroughly with 

Mi0004

 (tissue for wiping cuvets, see

chapter ACCESSORIES) or a lint-free cloth prior to insertion.

• Do not let the reacted sample stand too long after extraction,

or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is

important to take a new zero reading for each different wine-
sample. Use always a fresh prepared zero for making the zero
reading.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general

rule of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

Repeat step a), b) and c) until all the organic solvent is transferred. Ensure that at least 1/
3 of the cuvet is filled with organic solvent, otherwise erroneous results will be obtained.
Notes:
- If some wine sample is transferred too, this does not interferes with the method.
- Try to avoid transferring suspended solids that might be present.
- If necessary, air bubbles can be removed tapping the vials gently on the table.

Содержание Mi451

Страница 1: ...www milwaukeetesters com 1 www milwaukeetesters com Milwaukee Wine Lab Photometer Mi451 Copper INSTRUCTION MANUAL...

Страница 2: ...wine analysis 2 2 2 2 2 FUNCTIONAL DESCRIPTION 3 GENERAL DESCRIPTION 4 SPECIFICATIONS 6 GUIDE TO DISPLAY CODES 7 GENERALTIPSFORANACCURATEMEASUREMENT 8 MEASUREMENTPROCEDURE 10 ERRORMESSAGES 13 BATTERY...

Страница 3: ...R ON AND OFF E ZERO KEY TO START THE ZERO MEASUREMENT F READ KEY TO START THE SAMPLE MEASUREMENT G TIMER KEY TO ACTIVATE THE COUNTDOWN MODE TIMER DISPLAY A BATTERY STATUS ICON B HOURGLASS ICON C LAMP...

Страница 4: ...trations in wine result from contamination during post fermentation processing like contact with non stainless steel equipment and as impurities in fining agents and filter media The copper concentrat...

Страница 5: ...ncentrations limit SO2 addition high protein levels cold stabilize and bentonite fine to reduce proteins in white wine light and heat This meter is supplied with Two sample cuvets and caps Reagents fo...

Страница 6: ...filter 560 nm Method The reaction between Copper and the reagents causes a purple tint in the sample Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery Type 4 x 1 5 volt AA batteries A...

Страница 7: ...e timer is finished Indicates that the instrument is in a ready state and waiting for the next command Timer or Zero This prompt indicates the battery capacity value This prompt appears for a few seco...

Страница 8: ...mount of wine sample 15 mL or reagents A or B up to the 20 mL mark to the graduated glass vials The liquid in the vial forms a convexity on the top the bottom of this convexity must be at the same lev...

Страница 9: ...not let the reacted sample stand too long after extraction or accuracy will be lost It is possible to take multiple readings in a row but it is important to take a new zero reading for each different...

Страница 10: ...mL pipette to add 5 ml of Reagent B Mi551B 0 to the second vial B Add 4 full spoons of reagent C Mi551C 0 to each vial Using the other supplied spoon add 4 full spoons of Reagent D Mi551D 0 to each vi...

Страница 11: ...pette to transfer the upper layer organic phase into another empty cuvet If some wine is transferred too this does not interfere with the measurement Capthecuvet Thisisthereactedsample cuvet B Insert...

Страница 12: ...mg L ppm of copper on the Liquid Crystal Display Note If the copper concentration exceeds 1 50 ppm or if the sample is very turbid or dark red colored it is recommended to dilute the sample 10 times...

Страница 13: ...ess light than the zero reference Check the procedure and make sure you use the same cuvet for reference zero and measurement Inverted the sample and the zero cuvet are inverted ERRORMESSAGES The mete...

Страница 14: ...ets 2 pcs Mi0014 Caps for cuvets 2 pcs Mi0007 20 mL glass cylinder with caps 2 pcs BATTERYREPLACEMENT Battery replacement must only take place in a non hazardous area The blinking will appear when the...

Страница 15: ...user please return the intrument electrode and probe to either distributor or our office and the repair will be effected free of charge Damage due to accidents misuse tampering or lack of prescribed...

Страница 16: ...ls kik t sor 11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fa...

Отзывы: