Sterylis BASIC MANUAL
| 22
8
Possible errors and messages
All possible errors and messages are indicated by 4-digit codes on the display in the AUTOMATIC MODE section. If more
than 1 error occurs simultaneously, the codes are displayed one after the other, the presentation time of each code is 2
seconds. The meaning of all errors and messages is described in the table below.
Code
Description
E101
UV-C sources malfunction - incorrect power consumption
E102
Fan malfunction - incorrect power consumption
E201
Air filter error F1
E202
Air filter error F2
F001
Exceeded service life of the UV-C lamp
F002
Exceeded service life of air filters
F006
Air filters expiration date exceeded
Important notes on the occurrence of codes from the above table:
Exxx errors are stored and cannot be deleted by restarting the device.
Errors E201 and E202 should disappear when the filter cartridges are inserted back into the device again and
correctly.
Errors E101 and E102 can be reset by using the PRE-TEST and AUTO-TEST functions.
Fxxx messages can be deleted by resetting the corresponding counters:
F001 – counter d102
F002 – counter d105
F006 – counter d120