Sterylis  BASIC MANUAL                                                                                                                                                                            

  

| 15

 

 

During active front panel lock mode, press and hold the  UV-C and FILT buttons in the OPERATION TIME section for 3 
seconds simultaneously.  

6.5

 

Changing the configuration parameters 

To change the configuration parameters, press and hold the OK button in the "TIMER" section in the standby mode. Then 
enter the appropriate code to change the parameters. 
 

Entering the access code: 

The 4-digit code entered is shown on the display under AUTOMATIC MODE. The + and - buttons change the value by 1 
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The  OK 
button changes the active digit to the next one on the right. After entering the correct code, confirm it by long pressing the 
START button. If the code is incorrectly entered, the code HHHH will be displayed and a short beep will occur. In this case, 
the code must be re-entered. 
 
You can change the following parameters: 

Parameter 
code 

Description 

Min value 

Max value 

Default value 

C150 

Date setting - day of the month 

31 

15 

C151 

Date setting - month 

12 

C152 

Date setting - year 

2020 

2120 

2020 

C200 

FIRMWARE revision (read only) 

FW_REV 

 
The  individual  parameters  are  changed  using  the  +  and  -  buttons  in  the  AUTOMATIC MODE  section.  The  code  of  the 
selected  parameter  is  displayed  in  the  AUTOMATIC  MODE  section.  The  current  value  of  the  selected  parameter  is 
displayed in the OPERATION TIME section.

 

To change the value of the selected parameter, operate the + and - buttons in 

the OPERATION TIME section. The selection of the active digit is made by means of the RESET button. To confirm the 
value  of  the  selected  configuration  parameter,  press  and  hold  the  OK  button  in  the  AUTOMATIC  MODE  section.  The 
selected value will be confirmed by a short beep and all active displays will flash. 
 
To return to the standby mode, press and hold the OK button in the TIMER section. 

6.6 Working time counters 

The device is equipped with the operating time counters listed in the table below. The user has the possibility to read the 
indications of these counters and reset them.  
To view or reset the counters, press and hold the OK button in the TIMER section in the device standby mode. Then enter 
the appropriate code to change the parameters. 
 

Entering the access code: 

The 4-digit code entered is shown on the display under AUTOMATIC MODE. The + and - buttons change the value by 1 
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The  OK 
button changes the active digit to the next one on the right. After entering the correct code, confirm it by long pressing the 
START button. If the code is incorrectly entered, the code HHHH will be displayed and a short beep will occur. In this case, 
the code must be re-entered. 

 

You can interfere with the following counters: 

Parameter code 

Description 

d102 

UV-C lamp service interval timer [s] 

d105 

Fan service interval timer [s] 

d120 

Air filter interval timer [s] 

d201 

Active electricity consumption meter [kWh] 

Содержание STERYLIS BASIC-100

Страница 1: ...MODELS BASIC 100 BASIC 150 BASIC 300 BASIC 400 BASIC 800 BASIC 1200 USER S MANUAL PROFESSIONAL ROOM STERILISERS...

Страница 2: ...r disinfection standard mode operating mode 13 6 3 3 UV C air disinfection silent mode operating mode 13 6 3 4 UV C air disinfection automatic mode operating mode 14 6 4 Front panel locking mode 14 6...

Страница 3: ...information before you start servicing Before starting service work you should also read the UV C radiation safety data sheet All service work must be carried out in accordance with the instructions f...

Страница 4: ...manual please contact the manufacturer directly 7 It is essential that you read the meaning of the following warning icons They are located in the manual on the housing of the device or on other parts...

Страница 5: ...n pale cold skin sweat chills accelerated heart rate place them in an anti shock position on their back with their legs raised Despite a specially designed disinfection channel and limiting switch pro...

Страница 6: ...of more serious symptoms severe blinking pain foreign body impression in the eye it is necessary to contact an ophthalmologist Until an ophthalmologist has given assistance proceed as with lighter sy...

Страница 7: ...n drags the right amount of air through the filters and the UV lamp making the cleaning and disinfection process possible In filter mode the room air is cleaned with a pre filter F1 and an activated c...

Страница 8: ...W 45 W Average UV C radiation intensity 127 J m2 150 J m2 257 J m2 Net weight 16 kg 17 kg 31 kg IP degree of protection IP20 IP20 IP20 in UV C air disinfection standard mode Name STERYLIS STERYLIS STE...

Страница 9: ...obtain a high quality deodorising effect removing odours from the air Due to the special design of the UV C chamber the device operating in disinfection mode can be used in rooms where people are pre...

Страница 10: ...ditional 10 meter long power cord 4 Operating manual 5 Safety data sheet 6 UV C safety data sheet 7 Declarations of conformity in Polish and English if you purchase the device as an option with a long...

Страница 11: ...ned in the chapters Maintenance and Important safety instructions 6 1 Operation of the device To operate the room steriliser use the user panel on the front of the device The construction and operatio...

Страница 12: ...PERATION TIME section with operation time segmented display Alarm icons for fan or UV C fluorescent lamp system failure UV C section with graphic display User panel switch Operating modes section stan...

Страница 13: ...n process press and hold the START button located at the green filtration mode light again 6 3 2 UV C air disinfection standard mode operating mode To start the UV C air disinfection standard mode ope...

Страница 14: ...old down the button START in the AUTOMATIC MODE section The automatic mode light will start blinking the device will check the proper functioning of the relevant actuators PRE TEST function and then i...

Страница 15: ...ION TIME section To change the value of the selected parameter operate the and buttons in the OPERATION TIME section The selection of the active digit is made by means of the RESET button To confirm t...

Страница 16: ...ndby mode Then enter the appropriate code to move to the above mentioned functions Entering the access code The 4 digit code entered is shown on the display under AUTOMATIC MODE The and buttons change...

Страница 17: ...will go into standby mode 6 7 2 SET TIME function The function is used to set the current time on the controller When the SET TIME function is selected the number active on the display in the TIMER se...

Страница 18: ...s clog and avoid their excessive operation After the service life of the UV lamps 9 000 operating hours in the steriliser they should be replaced This is related not only to the possible burning of th...

Страница 19: ...nt air filter cartridge towards the floor 4 Remove the air filter from the cartridge 5 Insert a new air filter into the cartridge making sure that the direction of the arrow on the cartridge and on th...

Страница 20: ...in the UV C OPERATION SCHEDULE section If the limit violation ceases the error is automatically deleted The safety limiting switches provide protection in case of an attempt to change the air filters...

Страница 21: ...EDULE section If the limit violation ceases the error is automatically deleted The safety limiting switches provide protection in case of an attempt to change the air filters during operation 7 3 Repl...

Страница 22: ...on incorrect power consumption E201 Air filter error F1 E202 Air filter error F2 F001 Exceeded service life of the UV C lamp F002 Exceeded service life of air filters F006 Air filters expiration date...

Страница 23: ...sinfection mode in progress no active errors Green filtration mode light blinking tESt text blinking PRE TEST function for filtration mode in progress active errors Green silent mode light blinking tE...

Страница 24: ...for use within certain voltage limits Directive 2014 30 EU on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Directive 2011 65 EU on the restriction of th...

Страница 25: ...tment you may be required to provide the data on the unit s nameplate Serial number Date of manufacture Device name and model The information in this document may be changed by without notice Reproduc...

Отзывы: