Sterylis BASIC MANUAL
| 14
and then display the corresponding error codes on the display in the AUTOMATIC MODE section.
During operation of the device in disinfection mode, the display in the UV-C section is activated as standard.
The silent mode can also be directly selected if the device operates in any of the disinfection or filtration modes. To switch
from standard disinfection or filtration mode to silent disinfection mode, press and hold the START button located at the
green silent mode light.
To interrupt the silent disinfection process, press and hold the START button next to the green silent mode light again.
6.3.4 “UV-C air disinfection – automatic mode” operating mode
Automatic operation can be carried out in several programs: AUT.1, AUT.2, AUT.3 and AUT.4. The characteristics of these
work programs are presented in the table below.
Program code
Program name
Program description
AUT.1
WEEK
Work from 8.00 am to 4.00 pm, only on working days. Low fan speed
operation
AUT.2
HOME
Work from 6.00 pm to 9.00 pm, only on working days. Low fan speed
operation
AUT.3
MIX
Operation for 5 min at high fan speed and then 45 min at low fan speed
AUT.4
MAX_3H
3 hours of operation at high fan speed, then start silent disinfection mode
To start the unit in automatic disinfection mode, select the appropriate program code in standby mode. The program is
selected using the b and – in the AUTOMATIC MODE section. Then press and hold down the button START in the
AUTOMATIC MODE section. The automatic mode light will start blinking (the device will check the proper functioning of
the relevant actuators - PRE-TEST function) and then it will start to light steadily and the device will start working in the
selected operation mode.
If any errors are detected during the PRE-TEST function, the device will display a tESt message, enter the standby mode
and then display the corresponding error codes on the display in the AUTOMATIC MODE section.
During operation of the device in the automatic disinfection mode, the display in the UV-C section is activated as standard.
The automatic mode can also be directly selected if the device operates in any of the disinfection or filtration modes. To
switch from disinfection or filtration mode to automatic disinfection mode, press and hold the START button located at
the green standard mode light.
To interrupt the automatic disinfection process, press and hold the START button next to the green automatic mode light
again.
6.4 Front panel locking mode
The STERYLIS BASIC room sterilizer is equipped with the possibility of activating a user panel lock. It is designed to block
the device in such a way that unwanted people cannot operate it.
To activate the front panel locking mode:
During active standby, automatic disinfection, silent disinfection, standard disinfection or filtration, press and hold the UV-
C and FILT buttons in the OPERATION TIME section simultaneously for 3 seconds. Entering the lock mode will be
confirmed by displaying "L0CH" in the AUTOMATIC MODE section.
To deactivate the front panel locking mode: