Labscale Tangential Flow Filtration System User Guide
7
• Never operate the system without the tank cover in place. Never
operate the system if any moving parts are exposed. Never operate
the system with the pump housing open.
• Use appropriate personal protective equipment, including eye
protection, when operating the system.
This equipment can potentially be used with biohazardous
products.
Operating Requirements
The system needs to be connected to electrical power and have
fuses installed to operate.
The Labscale system must only be uesd with the Millipore wall
transfomer provided and Millipore accessories.
Location
Power Supply
Reference
North America
Japan
115 VAC 50/60Hz, 1-phase XX42LSS11
Europe,
China
230 VAC 50/60Hz, 1-phase XX42LSS12
United Kingdom 230 VAC 50/60Hz, 1-phase XX42LSS13
Operating Conditions
The maximum operating altitude
2000 m
The over voltage category
2
The temperature operating range
40 to 25 °C (39 to 77 °F)
in a non-condensing
environment
Storage Conditions
The temperature storage range
-10 C to 55 °C (14 to 131
°F) in a non-condensing
environment