Labscale Tangential Flow Filtration System User Guide
33
Drain System
See “Drain System” section, page 27.
Integrity Testing—Pellicon XL Device on System
Integrity Test Set Up
1. Close tank outlet valve and disconnect pump inlet tubing
(silicone, translucent) from the tank outlet valve. Close the
retentate valve.
Integrity Test Procedure
1. Obtain air integrity test pressure and airflow specifications
from the Certificate of Quality supplied with the device.
2. Insert syringe in the female luer end of the pump inlet tubing.
(silicone, translucent). Inject air into the tubing until the feed
pressure gauge reading reaches the recommended air integrity
test pressure.
Figure 32: Integrity Test Setup