PERMEATE TUBING
RETENTATE TUBING
Note: Check system connections for leaks—tighten connec-
tions if required
8. Adjust retentate valve restriction by slowly turning retentate
valve clockwise
until the retentate pressure gage reads
0.69 Bar (10 psi). Adjust pump speed and retentate valve
restriction to achieve 2.07 Bar (30 psi) feed pressure and 0.69
Bar (10 psi) retentate pressure.
9. Recirculate the cleaning solution for 30 - 60 minutes. Turn
pump off.