Labscale 500 mL Reservoir Setup
Note: All tubing lengths are recommended to minimize
recirculation volume. Longer lengths may be used.
After prolonged storage, the tubing may absorb a small
volume of water. As a result, the tubing color may
change from translucent to opaque, which is normal.
Air or oven drying will return the color to translucent.
Install Retentate Tubing
1. Cut silicone (translucent) tubing and install fittings as shown in
figure 10.
L = 15 cm (6 inches)
Fittings: 1 male, 1 female
MALE
FEMALE
L
L = Appropriate length to reach
collection vessel.
Fittings: 2 male
MALE
MALE
L
RET OUT PORT
RET IN PORT
RETENTATE TUBING
Figure 10:
Set-up of retentate tubing, fittings and clamps.
2. Remove plugs from retentate outlet (RET OUT) and retentate
inlet (RET IN) ports.
3. Insert the male luer end of the retentate tubing into the RET
OUT port and the female luer end of the retentate tubing into
the RET IN port. Turn fittings until snug.