Direct Detect® Spectrometer User Guide
33
After replacing the IR source, perform the Performance Qualification and
Operational Qualification instrument tests as follows:
11. Click on the instrument validation button
.
12. When the instrument validation dialog appears, click on the
Performance
Qualification Run Test
button (1 in Figure 36). The test takes about 5 minutes.
Figure 36.
Instrument validation dialog
2
1
13. After the test has finished, run the Operational Qualification test by clicking on
the
Operational Qualification Run Test
button (2).
14. If the instrument passes the test, the hardware diagnostics icon turns green.
If the instrument fails, refer to section 5.1 in the Direct Detect® Software User
Guide.
13.2 Replacing the Desiccant
Excessive humidity inside the spectrometer may cause H
2
O bands in the spectrum. To
reduce H
2
O bands to a minimum, the air inside the Direct Detect® instrument module is
convectively dried with a desiccant enclosed in a small, removable, multichambered pack.
If the desiccant is saturated and can no longer absorb humidity, it must be replaced.
Both the internal humidity icon (Figure 37) and instrument status light (on top of the
instrument, 5E in Figure 5) will turn yellow. When you click on the internal humidity icon,
a warning message with the current humidity level is displayed (Figure 37).