![Millipore DDHW00010-WW Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/millipore/ddhw00010-ww/ddhw00010-ww_user-manual_1796888030.webp)
Direct Detect® Spectrometer User Guide
26
12
Troubleshooting
12.1 General Information
The status of the Direct Detect® spectrometer is indicated by both the instrument
status light (located on the top right side of the instrument under the Netbook tray (5E
in Figure 5)) and the hardware diagnostics icon (located in the upper right corner of
the Direct Detect® Software user interface (Figure 30)). The status light and hardware
diagnostics icon colors are typically the same at any given time, and indicate the
following:
Green
Instrument is in proper operating condition (flashing green on the instrument
indicates stand-by mode)
Yellow
Instrument is initializing or indicating a warning (e.g., IR source has not yet
reached its operating temperature, the validity period of the instrument has
expired, the air humidity inside the spectrometer is too high)
Red
Instrument has malfunctioned, or has a defective component, or an OQ/PQ test
has expired or failed
Figure 30. Direct Detect® Software user interface
Instrument internal
humidity icon
Instrument validation icon
and OQ/PQ initiator
Hardware diagnostics icon
Содержание DDHW00010-WW
Страница 1: ...User Guide Direct Detect Spectrometer...