background image

M.I. P/N 004-2163 Rev. C 

3

loops that are an excellent illustration of the cardiac cycle in normal or diseased conditions 
of the living heart. 

With the MPVS 

Ultra

, P-V loops may be captured during pharmacological, therapeutic 

and hemodynamic interventions, allowing comprehensive, beat-by-beat evaluation of the 
fundamental mechanical properties of the heart. 

The MPVS 

Ultra

 supports the use of Millar’s ultra-miniature P-V catheters for small 

animal studies and multi-segment P-V catheters for large animal studies.  Both the small 
and large animal catheters couple a high-fidelity pressure sensor with volume electrodes 
spaced to match the longitudinal diameter of the left ventricle. The electrodes allow 
volume data to be acquired across a single segment of the left ventricle. 

The MPVS 

Ultra

 has an internal 20 kHz fixed-frequency oscillator, which generates either 

a 20µA or 100µA RMS sine-wave current that creates an electric field within the left 
ventricle of the heart. The system measures a corresponding voltage that is proportional to 
the electric field. The system continuously measures this voltage across the receiving 
electrodes. The measured voltage changes during the cardiac cycle as the left ventricular 
blood volume rises (diastolic filling phase) and falls (systolic ejection phase). 

The measured voltage decreases as volume increases (resistance of blood pool decreases, 
conductance of blood pool increases) and the measured voltage increases as volume 
decreases (resistance of blood increases, conductance of blood pool decreases). The system 
then converts measured voltages into ventricular conductance, which is proportional to the 
changes in ventricular volume as stated above. 

The MPVS 

Ultra

 also measures high fidelity pressure signals using a Millar catheter. The 

solid-state pressure sensor uses the same technology found in Millar’s extensive line of 
Mikro-Tip

®

 catheter pressure transducers. The use of the pressure sensor located between 

the conductance/volume electrodes allows the simultaneous measurement of pressure and 
conductance/volume, enabling a researcher to obtain high quality P-V loops from the 
rapidly beating hearts of small animals. The MPVS 

Ultra

 system should ONLY be used 

with a Millar catheter. 

 

Содержание MPVS Ultra

Страница 1: ...1 MPVS Ultra Signal Conditioning Hardware User Guide...

Страница 2: ...dditional guidelines 29 V Troubleshooting System Maintenance and Specifications 31 Hardware 33 Cables 34 Technical Specifications 35 Technical Support 38 Recommended accessories 39 Warranty 42 I Intro...

Страница 3: ...ge that is proportional to the electric field The system continuously measures this voltage across the receiving electrodes The measured voltage changes during the cardiac cycle as the left ventricula...

Страница 4: ...the equipment so that cables can easily be disconnected especially the power cord in case of an emergency This system must be connected to an easily accessible wall outlet that remains accessible afte...

Страница 5: ...th an ADInstruments PowerLab system should refer to the ADInstruments PowerLab setup section for specific instructions For users with other data acquisition systems packages of BNC to BNC cables are a...

Страница 6: ...d transducer will be connected to the P2 input on the front panel Use the Control Interface to select the input for the second pressure channel ADInstruments PowerLab Setup As shown below the MPVS Ult...

Страница 7: ...that the shield is connected to the ground pins indicated in the Theory of Operation section Users who build their own interface cables assume all responsibility for harm loss or damage related to th...

Страница 8: ...d in the procedure Place the cuvette under the animal or in a bath of body temperature water to warm the plastic This will minimize temperature effects on the resistivity rho measurement 2 Prepare the...

Страница 9: ...ata occlusion data stimulus reaction data and saline bolus data Smaller animals should have saline bolus maneuvers performed at the end of the study in case the saline induces detrimental effects on t...

Страница 10: ...appear Click on Load to set the displayed resistivity as the value to be used in calibration and measurements 3 Prepare the animal 4 Catheterize the subject Monitor the data acquisition system output...

Страница 11: ...ers with one or two pressure sensors When connecting to a three connector dual pressure P V transducer be sure to follow the labeling on the connectors to ensure that the pressure sensor inputs are no...

Страница 12: ...log voltage signal of this measurement is provided on this connection but users typically will not use this signal in connection with blood sample temperature With proper calibration a user can choose...

Страница 13: ...Segment capabilities a keycode will be supplied with the shipment For information about upgrading a MPVS Ultra Single Segment unit for use with Multi Segment PV catheters please contact Customer Servi...

Страница 14: ...will relock the features after sixty seconds have passed with no configuration changes By default the catheter configuration used when the program is opened will be the same as the configuration when...

Страница 15: ...Blood Resistivity This feature controls the resistivity used in Baan s equation for calculating the volume displayed on the gain and calibration controls within the Control Interface The values can al...

Страница 16: ...tation electrodes 4 There must be at least two active sensing electrodes The green boxes below the catheter graphic label the location of the segments that are being measured The labels S1 S7 correspo...

Страница 17: ...nductance Volume Range control are volume values or conductance values Create New Catheter Use this button to create a new entry in the Millar Catheter Selection table This is particularly useful if y...

Страница 18: ...nterface Pressure 1 Pressure 2 The Pressure 1 and Pressure 2 scales graphically display the currently measured pressure from 25 mmHg to 200 mmHg This number is also numerically displayed The calculati...

Страница 19: ...ressure as zero Pressure 2 Select The MPVS Ultra provides two amplifiers for use with a pressure transducer Pressure 1 is always set to the P V input connector on the MPVS Ultra front panel This will...

Страница 20: ...ns are available at any given gain range Available buttons are beige others are gray The available selections are designed to span the expected segmental conductance inputs for typical animals ranging...

Страница 21: ...To turn this filter on select On and to turn it off select Off ECG Input The ECG Input allows the user to choose the source of the ECG input signal The MPVS Ultra can record an internal ECG signal fro...

Страница 22: ...number written on the cable Enter the K factor for your system The gain cannot be changed in this view Click OK then click on the new entry in the Millar Cuvette Selection table Delete Highlighted Cuv...

Страница 23: ...urement F C 4 With as little delay as possible do steps 4 6 Place the cuvette on a flat surface and dry the well with a gauze or swab 5 In a heparinized syringe draw about 3cc to 5cc of blood Place th...

Страница 24: ...cuvette should always be emptied and cleaned immediately after use to prevent blood from clotting within the cuvette Procedure for cleaning the cuvette 1 Properly dispose of the used blood sample 2 R...

Страница 25: ...ventricular volume changes can be simplified in terms of Ohm s Law Ohm s Law V IR G 1 therefore R 1 x R G V I therefore G I x G V Volume G I x V ____________________ V voltage potential between E2 and...

Страница 26: ...ion current through a selected primary excitation electrode E5 through E11 located at the proximal end of the measurement area and through primary excitation electrode E1 located at the distal end of...

Страница 27: ...y is a material property whereas resistance and therefore conductance is a property of a sample or a structure When measuring the conductance of the blood in the ventricle the reading is based on the...

Страница 28: ...will be reduced and the signal will be ideal for timing in PVAN Ultra calculations The input to the IECG signal is the electrodes that are presently in use for the primary excitation field This ensure...

Страница 29: ...the inputs to the thermistor Do not connect to pins 1 2 3 or 4 since doing so may damage the resistivity measurement circuit Pin 7 may be used to ground the shield of the cable It is very important to...

Страница 30: ...e and conductance Conductivity is the multiplicative inverse of resistivity Conductance is the multiplicative inverse of resistance G 1 R Conductance and conductivity relate to how well a structure or...

Страница 31: ...ter tip Do not submerge the catheter s electrical connectors as doing so will damage the sensitive electronics housed inside The presoak helps prepare the pressure sensor diaphragm for the wet biologi...

Страница 32: ...s to grip the Mikro Tip P V catheter slip some soft tubing over the tips of the metal instrument to cushion the interface between the instrument and the catheter body as shown in the photo Always be m...

Страница 33: ...s disconnected from the monitoring equipment The Mikro Tip P V catheter should be thoroughly inspected under magnification after each use to determine its condition Carefully examine the catheter for...

Страница 34: ...onnectors Contact Millar or your distributor for service Cables The cables can be wiped with isopropyl alcohol 70 Do not submerge the cables or cords When connecting and disconnecting the cables and c...

Страница 35: ...nominal Input connector for system Five pin Redel or ODU Pressure Outputs Specification Value Sensitivity 1 V 100 mmHg nominal Accuracy Error Band 0 200mmHg 1 mmHg or 1 of reading whichever is greate...

Страница 36: ...e 10 VDC Noise 75mVPP Input Connector 27 pin Alden Pulse Lok Pulse Lok is a registered trademark of Amphenol Alden Products Company Inc Output Connectors BNC connector Composite S1 S2 S3 S4 S5 S6 S7 V...

Страница 37: ...resistance to power cord ground pin 7 ft cord 1 ohm Enclosure Specification Value Ingress protection IP 20 Dimensions 12 3 8 x 11 3 4 x 2 7 8 width x length x height 31 4cm x 29 7cm x 7 3cm width x l...

Страница 38: ...r Off System power Technical Support Technical Support The Purchaser is entitled to free technical support for any Millar product Millar s technical support staff can provide help and advice concernin...

Страница 39: ...were shipped with the system Use these part numbers when ordering 850 5105 North American cord with NEMA 5 15 plug 2m 6 7 long 850 5118European cord with CEE 7 VII plug 2 5m long 850 5117North America...

Страница 40: ...dard BNC cables are required for connecting to the outputs of the MPVS Ultra BNC cable quality is important for ensuring clean signals so be sure to use coax cables with a braided copper shield Millar...

Страница 41: ...M I P N 004 2163 Rev C 41 Connector part number PAGM07GLAC65GZ Strain relief part number GMA 1B 054 DG Rho Cuvette 910 1060Rho Cuvette Technical Specifications...

Страница 42: ...other conditions cited in this Agreement General Limitations Millar products are produced to high standards and should perform substantially as described in the supplied documentation There is a limit...

Страница 43: ...But Not Manufactured by Millar Millar supplied products of a third party manufacturer i e data acquisition systems amplifiers computers monitors and printers including associated cables probes accesso...

Страница 44: ...ser via the following options Phone 832 667 7000 8 00AM 5 00PM Central Time M F Fax 713 714 8497 24 hours a day 7 days a week Email info millar com 24 hours a day 7 days a week Web millar com On site...

Отзывы: