M.I. P/N 004-2163 Rev. C
23
2.
In the
Millar Cuvette Selection
field, select the appropriate configuration. By default,
the software will use the configuration that was used most recently. The configuration
includes both the gain and the K factor.
3.
Select the desired units for temperature measurement (
°F
/
°C
).
4.
With as little delay as possible, do steps 4-6. Place the cuvette on a flat surface and dry
the well with a gauze or swab.
5.
In a heparinized syringe, draw about 3cc to 5cc of blood. Place the tip of the needle at
the bottom of the well and inject the blood into the well. Ideally, the surface of the blood
should be level with the top surface of the cuvette.
6.
Click on
Measure
in the Control Interface. The
Measured Snapshot
field
will appear.
Detailed Operation Detailed Operation