Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
79
3.7.4
Modify a WAX cleaning
Select Cleaning from the MAIN FUNCTIONS screen, then press Modify Cleaning Program.
The screen on the right appears, showing the existing cleaning programs for the MAIN CAVITY(on the left)
and for the WAX (on the right): default protocols as well as protocols created by the USER (if any).
Choose the program of your interest (figure below, arrow). Press Copy button (figure below, rectangle), re-
name the program and modify the fields.
Take care to modify ONLY the Time At Temperature part of the curve.
Do not modify the Ramp - up time.
The Modify and Delete buttons are made available only for already copied protocols.
For safety reasons, the factory pre-stored protocols cannot be directly modified or
deleted.
The Modify and Delete buttons are made available only for already copied protocols.
Take care to modify ONLY the Time At Temperature part of the curve.
Do not modify the Ramp - up time.