Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
104
7.3 Every 5 runs
It’s required to clean the wax in order to extend its reuse life and avoid artifacts related to the impregnation
step, as described in the chapter 3.5.
Wax cleaning takes 90 minutes.
7.4 Weekly
Check if there is any residual reagent in the drawer. If you want to empty it, unscrew the cup as indicated in
the figure below, put a container under the unit and unload the residual reagent.
CLEANING PROCEDURE FOR CLEANING TANKS
TO BE PERFORMED AFTER EVERY CHANGE OF CLEANING REAGENTS
Reagents required:
Xylene, Isoparaffin or substitute and Ethanol.
Purpose of cleaning:
is performed on all CLEANING tanks that have potential contamination from WAX
residues.
When to clean:
to be performed EVERY TIME cleaning reagents are changed.
Removal of tanks:
remove the pipe and then the tank. Dispose of reagent according to lab practices
appropriate for that reagent.
Cleaning procedure:
perform a rinse in Xylene or Isoparaffin or substitute (close tank and shake for a
few minutes), dispose of waste reagents and follow with two quick rinses in Ethanol.
Reinstalling of tanks and reagent replacement:
load tanks with fresh cleaning reagents, put the pipe
and then the tank in the drawer. Complete replacement of reagent that resets reagent management
counts.
It’s suggested to clean the every 5 runs in case LOGOS ONE is working once a day
without sponges. In any other case, like use of sponges or several runs per day, please
contact our Application Dept at
to get proper
instructions.
Wax must be completely replaced (and NOT cleaned) at least twice a month, depending on
the use of LOGOS ONE. See the chapter 4.3 for the wax replacement procedure.
It’s required to execute the cleaning procedure when the cleaning tanks are not
commercial but either supplied locally or purchased directly from the manufacturer
because it’s possible to reuse them several times.