– LOGOS ONE – Operator Manual
42
GENERAL
Select General to modify:
Institute and Department
name where unit is installed.
Reagent Cassettes Alert Status
: this number indicates that the reagent status is near expiration (on the
base of the number of cassettes processed) and the reagents status icon turns from green to yellow (see
chapter 4).
Reagent Cycles Alert Status
: this number indicates that the reagent status is near expiration (on the base
of the cycles run) and the reagents status icon turns from green to yellow (see chapter 4).
Reagent Days Alert Status
: this number indicates that the reagent status is near expiration (on the base of
the days of use) and the reagents status icon turns from green to yellow (see chapter 4).
VENTILATIONS
Select Ventilations to manage exhaust fumes fans of cavity (working area) and tanks (holder), choose either:
Always active
: both exhaust fans are activated at the switching ON of the unit.
Always off
: both exhaust fans are always OFF, even during the processing.
NOT SUGGESTED.
Risk of fumes in the laboratory environment.
Processing and cover opening
: the cavity ventilation is activated only during the processing and when
opening the covers while the tanks ventilation is always active.
REMOTE CONNECTION
Select
Remote connection
to edit and modify the network settings for remote assistance.
Data required must be supplied by the Laboratory Network ADMINISTRATOR (Primary WINS not required).
FIXATION LIMITS
Select Fixation limits to set Minimum and Maximum Fixation time. When enabled, before starting a protocol,
the software warns USER if the total fixation time is under or over the limits set and manages the Holding
solution. Refer also to chapter 3.2.5 and 4.1.1.
RUN ID
If Run ID is ticked, software requires user to enter the Run ID before running a program.
For further details see chapter 2.8.
Enable or disable Run ID check according to your internal laboratory practice
(Good Laboratory Practice, GLP).