Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
58
2.
On completion of the UNLOADING (and of vaporization, if required by the program in use), it is necessary
to MANUALLY move the RACK from the MAIN CAVITY to the WAX CAVITY. A message with an audio
alarm warns the operator:
A.
If operator does not press OK within 3 minutes, WAX is automatically loaded in the MAIN CAVITY.
Removal of the rack must be done within 3 minutes.
Should this not be done, the system AUTOMATICALLY pumps WAX in the MAIN CAVITY to
preserve samples.
If the RUN MODE becomes AUTOMATIC, after it is required to run a MAIN CAVITY
CLEANING.
B.
If operator presses OK this message appears with an audio alarm:
A.
If operator presses OK, WAX IMPREGNATION starts immediately IN WAX CAVITY.
B.
If operator does not press OK, after 1 minute WAX IMPREGNATION starts IN WAX CAVITY.
NOW MOVE MANUALLY THE RACK IN THE WAX CAVITY (right).