Guidance and manufacturer’s declaration - electromagnetic immunity
Mains-frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m,
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typ-
ical commercial or hos-
pital environment.
Electrical fast transients/
bursts
IEC 61000-4-4
Input a.c. power port:
±2 kV
Input d.c. power port:
±2 kV
Signal input/output parts
port: ±1 kV
(with 100 kHz repetition
frequency)
Input a.c. power port:
±2 kV
Input d.c. power port:
±2 kV
Signal input/output parts
port: ±1 kV
(with 100 kHz repetition
frequency)
Mains power quality
should equal that of
usual commercial or hos-
pital environments.
Note: The miha
bodytec m.ove EMS
training device has no
input d.c. port and no
signal input/output parts
ports with a maximum
cable length of 3 meters
or more.
Guidance and manufacturer’s declaration - electromagnetic immunity
The miha bodytec m.ove EMS training device has been designed for use in electromagnetic environments
as outlined below.
The buyer or user of the miha bodytec m.ove device shall make sure that the device is used under such
conditions.
Immunity test
IEC-60601 test level
Compliance level
Electromagnetic envi-
ronment - guidance
Surges
IEC 61000-4-5
Input a.c. power port:
±0.5 kV, ±1 kV, ±2 kV
line-to-ground ±0.5 kV,
±1 kV, line to line
Input d.c. power port:
±0.5 kV, ±1 kV, ±2 kV
line-to-ground ±0.5 kV,
±1 kV, line to line
Signal output parts port:
±2 kV
Input a.c. power port:
±0.5 kV, ±1 kV, ±2 kV
line-to-ground ±0.5 kV,
±1 kV, line to line
Input d.c. power port:
±0.5 kV, ±1 kV, ±2 kV
line-to-ground ±0.5 kV,
±1 kV, line to line
Signal output parts port:
±2 kV
Mains power quality
should equal that of
usual commercial or hos-
pital environments
Electromagnetic compatibility
2021-07-28
miha-bodytec m.ove EMS training device