Guidance and manufacturer’s declaration - electromagnetic immunity
The miha bodytec m.ove EMS training device has been designed for use in electromagnetic environments
as outlined below.
The buyer or user of the miha bodytec m.ove device shall make sure that the device is used under such
conditions.
Immunity test
IEC-60601 test level
Compliance level
Electromagnetic envi-
ronment - guidance
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz mod-
ulation 80% AM at 1 kHz
10 V/m
The minimum separation
distance d in m to a RF
emitting device with a
maximum power P in W
not operating in a fre-
quency band covered by
the proximity fields test
can be calculated using
the equation in clause
8.10 of IEC 60601-1-2
with E as the field
strength in V/m taken
from the compliance
level:
E = 6 / d ∙ √P
Proximity fields from RF
wireless communications
equipment IEC
61000-4-3
385 MHz: 27 V/m
450 MHz: 28 V/m
710 MHz: 9 V/m
745 MHz: 9 V/m
780 MHz: 9 V/m
810 MHz: 28 V/m
870 MHz: 28 V/m
930 MHz: 28 V/m
1720 MHz: 28 V/m
1845 MHz: 28 V/m
1970 MHz: 28 V/m
2450 MHz: 28 V/m
5240 MHz: 9 V/m
5500 MHz: 9 V/m
5785 MHz: 9 V/m
385 MHz: 27 V/m
450 MHz: 28 V/m
710 MHz: 9 V/m
745 MHz: 9 V/m
780 MHz: 9 V/m
810 MHz: 28 V/m
870 MHz: 28 V/m
930 MHz: 28 V/m
1720 MHz: 28 V/m
1845 MHz: 28 V/m
1970 MHz: 28 V/m
2450 MHz: 28 V/m
5240 MHz: 9 V/m
5500 MHz: 9 V/m
5785 MHz: 9 V/m
Portable RF communica-
tions equipment
(including peripherals
such as antenna cables
and external antennas)
shall be used no closer
than 30 cm (12 inches)
to any part of the miha-
bodytec m.ove EMS
training device, including
cables specified by the
manufacturer. The field
strength test levels are
calculated using the
equation in clause 8.10
of IEC 60601-1-2:
E = 6 / d ∙ √P with E as
the field strength in V/m,
the maximum power P in
W of and the minimum
separation distance d in
m to the RF.
Electromagnetic compatibility
2021-07-28
miha-bodytec m.ove EMS training device
43