Operation
60
Cleaning the filter
combination and
tubular filter
The filters in the wash cabinet must be checked daily and cleaned
regularly, see “Maintenance/Cleaning the filters in the wash cabinet”.
The A 800 tubular filter can be used in special injector manifolds on
various mobile units and baskets and must be cleaned regularly.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
A counter in the controls can be activated to remind you of the
required cleaning at regular intervals.
Clean tubular filter?
Clean filter combination?
Now
Later ( Programmes)
Now
Later ( Programmes)
Follow the instructions in the “Maintenance” section for cleaning the
filters. For the tubular filter, follow the instructions in the operating
instructions for the tubular filter.
Resetting the
counter
The counter for the cleaning interval may be reset only after
cleaning has been completed.
Tubular filter cleaned?
Filter comb. cleaned?
Yes
No
Yes
No
-
Yes
The counter is reset.
-
No
The counter will not be reset.
Select an option using the
and
arrow buttons and confirm your
choice with
OK
.
Low fill levels
When the fill level is low in one of the containers, for example for
process chemicals or the salt container, you are prompted to refill the
container.
Refill salt
DOS Refill
Acknowledge the messages with
OK
and refill the relevant
containers. To do this, follow the instructions in the sections “Filling
process chemicals and dispensing” and “Water softener”.