![Miele PG 8591 Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/miele/pg-8591/pg-8591_operating-instructions-manual_237375036.webp)
Areas of application
36
Wash pressure measurement
The wash pressure can be measured on all mobile units and baskets
with spray arms, injector manifolds or other wash connections, e.g.
during performance tests and validations in accordance with EN ISO
15883.
Test point for
measuring wash
pressure
On mobile units and baskets with spray arms and additional injector
manifolds or other wash connections, there is a connection on the
injector manifold or a wash connection for wash pressure
measurement. The exact location is described in the respective
operating instructions for the mobile units and baskets.
On mobile units and baskets with spray arms and no additional wash
connections, the test point for measuring wash pressure is provided
on the side of the water supply pipe.
Under no circumstances may load items, irrigation connectors etc.
be connected to the test point. After the measurement, the test
point must be closed again with the blind stopper.