Connecting the H
2
O
2
solution
container
The connection hose for the H
2
O
2
solution container is identified by a
black label. The connection hose is
supplied without a safety cap as the
container for hydrogen peroxide will
vary depending on supplier.
^
Connect the supplier's safety cap to
the connection hose (black).
^
Then connect this to the H
2
O
2
solution container.
^
Then start the
DOS2-FILL
programme.
Unlike with other chemical agents the
H
2
O
2
solution container must be fully
emptied before it is exchanged for a
new one.
^
Do not replace the H
2
O
2
solution
container with a new one until the
Fill
DOS2 container
message is
displayed. After connecting the new
container, you must run the
DOS2-FILL
programme.
^
If the
Check dispensing system 2
message appears you should check
the container as well as the
dispensing system. The programme
will stop automatically.
Special processes - OXIVARIO and ORTHOVARIO
79