DI water connection
The machine is equipped for optional
connection to a DI water supply. When
the water pressure is below the
recommended pressure, the water
intake time is automatically extended.
Acceptable water pressure (flow rate):
10 - 147 psi
Recommended water pressure (flow rate):
25 - 60 psi
^
Connect the DI hose (pressure-tested
and marked with "H
2
O pure") with 3/4"
GHT male thread to the on-site water
valve for distilled water.
If the DI connection will not be used,
the electronics may need to be
reprogrammed. The water intake
hose then remains at the rear side of
the machine unconnected.
The DI water connection should not be
hooked up (or Y connected) to any
water source, it should remain
unconnected, unless DI water is
available. The instructions below
re-program the unit to draw cold tap
water for the final rinse instead of the DI
water.
Access programming mode by
pressing and holding the "Display" and
"Start" buttons while simultaneously
turning the machine on.
^
Press "Drying" once and E01
(programming level) will appear in
the display.
^
Turn the program knob to the
11 o'clock position (will most likely
display 110).
^
110 = Set for DI water intake for final
rinse.
^
111 = Always cold water for final
rinse instead of DI water.
^
Press "Display" until 111 appears in
the display.
^
Press "Start" twice to save.
Plumbing