26-0002-0005/0 (en) (2022)
57
The Wireless Transmitter is intended for use in the radiated electromagnetic fields specified below. The
customer or the user of the Wireless Transmitter should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
Level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Wireless
Transmitter, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to
80 MHz
𝑑𝑑
= 1.2
√𝑃𝑃
Radiated RF
IEC61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to
2.5 GHz
𝑑𝑑
= 1.2
√𝑃𝑃
80 MHz to 800 MHz
𝑑𝑑
= 2.3
√𝑃𝑃
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strength from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all
situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and
people.
a
Field strength from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio
broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field
strength in the location in which the Wireless
Transmitter is used exceeds the applicable RF
compliance level above, the Wireless Transmitter
should be observed to verify normal operation. If
abnormal performance is observed, additional
measures may be necessary, such as re-orienting or
relocating the Wireless Transmitter.
b
Over the frequency range 150 kHz to 80 MHz, field
strength should be less than 3 V/m.
Table 17.3: Guidance and Manufacturer’s Declaration—Radiated Electromagnetic Fields
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...