26-0002-0005/0 (en) (2022)
49
•
Software restrictions
—No other software shall be loaded onto the SAPS laptop, because it
may interfere with proper functioning of the SAPS already installed on the commercially
available laptop.
•
Laptop use
—Using the commercial laptop containing SAPS for other purposes (such as
internet access) may damage the system or interfere with proper functioning of the SAPS.
•
Use only SAPS software
—The use of any software other than that specified herein will
violate safety, effectiveness, and design controls of the programmable devices, and such
use may result in an increased risk to users and patients.
•
Interference information
—This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses, and can radiate RF energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try and correct the interference by one or more of the following measures:
−
Reorient or relocate the receiving antenna.
−
Increase the separation between the equipment and receiver.
−
Connect the equipment to an outlet on a circuit different from that to
which the receiver is connected.
−
Consult the dealer or an experienced radio/TV technician for help.
•
EMI interference
—Programming in the presence of an EMI signal can be difficult or
impossible, but problems can usually be resolved by repositioning the patient, the WT, or
the EMI source.
•
Avoid large metal structures when programming
—Programming while the IPG and/or
patient are on a metal table or near large metal structures can be difficult or impossible, but
problems can usually be resolved by placing the WT closer to the IPG or moving to a
different location.
•
No internet
—Internet access has been purposely disabled on the laptop. Do not enable
the internet on the SAPS laptop.
•
Do not modify equipment
—Do not modify this equipment without the authorization of
MicroTransponder. Any modifications to the unit, unless expressly approved by
MicroTransponder
,
could void your authority to operate this equipment.
•
Set laptop time
—It is important that the laptop time be set to the correct local time and
periodically verified. This time is used to set the device time and to timestamp the
operations in the log file.
•
The magnet effects on other devices
—The magnet provided to the patient to start and
stop at-home therapy may damage televisions, computer disks, credit cards, and other
items affected by strong magnetic fields. The patient should keep the magnet away from
such items.
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...