26-0002-0005/0 (en) (2022)
24
Figure 4.16: Stim Test Mode
Figure 4.17: Implant Settings
1. Select the button next to
Stim Test Mode.
The button will now toggle to
On
(see Figure
2. Optionally, you may change the stimulation parameters displayed in the
Implant Settings
menu (see Figure 4.17). For each of the settings (
Amplitude
,
Frequency
,
Train
Duration
,
Pulse Width
), click the button next to the setting, and a menu will drop down
containing all available parameters for that setting. Then select the appropriate value for