Once a session is initiated, you may trigger stimulation, implementing the programmed IPG
therapy parameters. To stimulate, use the mouse and cursor to select the
Stimulate
button or
press the push button.
When the stimulation has been triggered, the
Status
window displays “Stimulating” for the
length of the stimulation period (see Figure 5.). Meanwhile, the
Stimulate
button is disabled for
double the time of the stimulation train duration (to allow for a 50% duty cycle). Stimulation can
occur only as often as twice the stimulation train duration; no duty cycle over 50% is allowed by
the software. For example, if a 0.5 second train duration is given, you cannot activate another
stimulation for at least 0.5 seconds after the train duration ends.
Figure 5.5: Stimulate Selected—Status “Stimulating”
The message “Refractory” appears in the
Status
window for a length of time equal to the
programmed stimulation train duration immediately following the “Stimulating” message (see
Figure 5.).
Figure 5.6: Status “Refractory”—No Stimulation Allowed
The value in the
Stimulation Counter
window increases by one for every stimulation delivered.
This counter resets to 0 when the current task is ended and a new task started (see Figure 5.).
Figure 5.7: Stimulation Counter
Once the stimulation ends and the refractory period has finished, the message in the
Status
window returns to “Ready to Stimulate”, the
Stimulate
button is once again selectable, and a
new repetition of the task can be initiated and stimulated (see Figure 5.).
The user may then continue to deliver stimulation except during the stimulation refractory
period.
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...