Configuration Database
MS-860 Industrial Raster Reader User’s Manual
5-3
Setup for Configuration Database
When setting up the database, the following parameters can be changed:
Definition:
When setting up the configuration database, for each index number (1 to
10) the values for 11 parameters can be defined applied during the read
cycle as enabled by
"Number of Active Indexes"
on page 5-4.
Serial Cmd:
<
K253,index
,gain,AGC mode,tracking,raster top,raster bottom,raster
speed,laser framing status,laser on position,laser off position,laser
power,background color>
Default:
0
(Index)
Options:
1 to 10 indexes, 0 = disabled
Parameter
Also in
Calibration?
a
a. It is important to note that the above settings are not identical to the settings changed by the cali-
bration routine described in Chapter 4, “Calibration.”
Reference
Gain
Yes
See
<K504>
,
"Gain Level"
on page 7-17
AGC mode
No
See
<K504>
,
"Automatic Gain Control (AGC)"
on page 7-19
Tracking
Yes
See
<K520>
,
"Tracking"
on page 7-17
Raster Top
Yes
See
<K506>
,
"Top Offset"
on page 8-4
Raster Bottom
Yes
See
<K506>
,
"Bottom Offset"
on page 8-4
Raster Speed
Yes
See
<K506>
,
"Raster Sweep Rate"
on page 8-5
Laser Framing Status
No
See
<K700>
,
"Laser Framing"
on page 8-7
Laser On Position
Yes
See
<K700>
,
"Laser On Position"
on page 8-8
Laser Off Position
Yes
See
<K700>
,
"Laser Off Position"
on page 8-8
Laser Power
Yes
See
<K700>
,
"Laser Power"
on page 8-8
Background Color
No
See
<K451>
,
"Background Color"
on page 9-31
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com