4. PRECAUTIONS FOR USE AND MAINTENANCE
STORAGE AND TRANSPORTATION
The Spider SAS is packaged as outlined in the packaging chapter at time of delivery.
The following table shows the conditions of storage and transportation for Spider SAS:
Storage and transportation
Temperature
Humidity
Pressure
Recorder
-20°
~
65°C
5%
~
85%
non condensing
500 hpa
~
1060 hpa
μSD card
-20°
~
65°C
8%
~
95%
non condensing
500 hpa
~
1060 hpa
Electrode
5°
~
30°C
NA
NA
Cable
10°
~
40°C
25%
~
90%
non condensing
NA
A protective bag is needed in case of extremely humid environment.
Remove the battery before storing Spider SAS.
CAUTION:
If stored over a prolonged period,
remove the battery
from Spider SAS.
According to the duration and conditions of storage and the battery quality type, there is a
risk of electrolyte leak
into the Spider SAS’s plastic case. In such a case the device must
be returned to the MicroPort After Sales Department for repair. Repair is excluded from the
warranty and a fee will be charged.
CAUTION
: If a leak or a powdery deposit is visible on the battery, in the compartment or on
the electrical contacts, remove the battery and dispose of it following local environmental
legislation and/or the intended procedures.
If the leak is not significant (no oxidation on
the electrical contacts of the battery), it is possible to clean it with a damp cloth, and
then carefully dry the cleaned parts
. In the event of traces of significant oxidation, it is
strongly recommended not to use the device (risk of recording failure) and return it to
MicroPort After Sales. Repair is excluded from the warranty and a fee will be charged.
CONDITIONS FOR USE
CAUTION:
If a patient carrying a Spider SAS recorder needs to be defibrillated, it is
mandatory that the recorder’s electrodes be removed before using the defibrillator. Spider
SAS is not protected from the effects of external defibrillation shocks or against high
frequency signals or strong electromagnetic disturbances.
The table below illustrates conditions for use for Spider SAS
Use
Temperature
Humidity
Pressure
Recorder and
µSD card
0°
~
38°C
10%
~
95%
no condensing
700 hpa
~
1060 hpa
The correct operation of Spider SAS is only guaranteed when used with µSD cards provided
by MicroPort.
4.1.
4.2.
4. PRECAUTIONS FOR USE AND MAINTENANCE
12
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F