CAUTION
: Spider SAS does not start if a MicroPort SD card is not inserted.
NOTE:
µSD cards designed for MicroPort devices cannot be used with other devices.
µSD cards can be used up to 300 000 times. These cards resist electrostatic discharges of
± 4kV maximum in the event of direct contact with the electrodes and ±8 kV apart from the
zone of the electrodes (flat contact). The µSD cards can resist a shock of 1 000 G maximum.
CAUTION
: Spider SAS shall not be used in environments with flammable anesthetic
mixtures containing oxygen or nitrous oxide.
ELECTROMAGNETIC ENVIRONMENT
The use of portable and mobile RF communication equipment (e.g. cellular phones) can
affect the Spider SAS since the recorded signals may be disturbed due to electromagnetic
interference.
CAUTION
:
The device shall not be used
in the presence of ionizing radiations (x-rays,
gamma rays...) or MRI System that could damage Spider SAS, φ Hub and/or µSD card.
Remove sensors and devices before starting any MRI procedure.
Concerning the electromagnetic emissions and the level of the immunity of the recorder, see
Appendix 2: Electromagnetic compatibility.
CARE AND REPAIR OF THE DEVICE
No particular preventive maintenance is required for Spider SAS, Physio Hub and their
accessories.
However, should the device fail to work properly or if plastic enclosure is broken after
mechanical shock for example, contact your local MicroPort affiliate, or the MicroPort After-
Sales Service. Do not unscrew the plastic case of the device under any circumstances and
do not attempt to repair or modify mechanical or electrical parts since this may lead to
malfunctions.
WARRANTY
The Spider SAS, Physio Hub and µSD card are guaranteed for two years from the date of
delivery. After-Sales Service shall repair any faulty device within the warranty condition free
of charge. After-Sales Service shall still be available after the warranty has expired but with a
fee for repair.
All reusable sensors, in this case, oximeter and piezo sensors, are guaranteed for 1 year.
All reusable accessories are guaranteed for 3 months including ECG cables, Bluetooth
dongle, bags, pouches, blue and beige belts.
The 1-year warranty applies neither to consumables nor to single-use devices.
NOTE:
The warranty is only valid under the condition that no attempts have been made to
open or repair the device. The warranty shall be void if devices or cards have been used
contrary to this user manual’s recommendation.
Only µSD cards provided by MicroPort are under guarantee.
4.3.
4.4.
4.5.
4. PRECAUTIONS FOR USE AND MAINTENANCE
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F