5. CLEANING
CAUTION:
A recently cleaned device must be dried carefully to avoid any oxidation of the
electrical contacts before long term storage.
SPIDER SAS AND PHYSIO HUB
The external case and the patient cable (including snap-on clips) must be cleaned with
professional liquid soap and disinfectants.
Do not use detergents, alcohol or acetone.
The electrical parts and the connectors should be cleaned daily in order to avoid dysfunctions.
After being cleaned, the materials must be carefully dried before storage and use. Spider SAS
and Physio Hub are not waterproof devices, therefore they should not be immersed in liquid.
CAUTION:
Do not immerse the snap-on cable clips in a cleaning or disinfectant solution.
This method considerably shortens the life of cables and causes an important internal
oxidation between the clip and the wire of the cable. The quality of the ECG signal could be
affected.
SPO
2
SENSOR
Reusable
PureLight®
Sensor
is
manufactured by Nonin® Medical Inc.
Cleaning instructions are available in the
"Instructions for use" included with the sensor
or on the Nonin website
(http://www.nonin.com/ProductCleaning).
CAUTION:
Do not submerge or immerse sensor or connector plug. Do not place device in
liquid or clean it with agents containing ammonium chloride or products that are not listed in
the User’s Guide supplied by Nonin Inc.
Clean reusable sensors between patients or as needed:
1. Use a soft cloth with a 10% bleach/90% water solution (household bleach < 10% sodium
hypochlorite).
2. Wipe away from connector plug. Do not turn Soft Sensors inside out.
3. Leave the Soft Sensors to dry completely before next use
5.1.
5.2.
5. CLEANING
16
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F