Cause A:
Power cord not fully inserted at one end.
Action A:
Insert power plug firmly into power source and analyzer power connecter.
Cause B:
No power at outlet.
Action B:
Plug in another electrical device to test the outlet. If there is no power, contact
an electrician.
Cause C:
Plug prongs bent so that contact not made at outlet.
Action C:
Gently move power plug at outlet while watching power indicator light. If the
indicator light comes on, have electrician replace outlet or plug.
Cause D:
Power cord damaged.
Action D:
Have electrician check cord using a test meter. Replace the cord if defective.
Cause E:
Loose internal connection or broken wire.
Action E:
Call a Micromeritics Service Representative for repair or replacement inform-
ation.
Valves cannot be operated.
Cause:
Cable from computer to the instrument is loose.
Action:
Reconnect the cable.
Vacuum pump is noisy.
Cause A:
Sample tube connector is loose.
Action A:
Tighten fitting. Replace O-ring.
Cause B:
Sample tube O-ring is worn or cracked.
Action B:
Replace O-ring. See
Sample Tube O-ring Replacement on page 10 - 25
.
Cause C:
Sample tube is cracked.
Action C:
Replace with new sample tube.
Cause D:
No sample tube loaded on a selected port.
Action D:
Install plug or empty sample tube.
Cause E:
Gas inlet valve open while vacuum valve open.
Action E:
With manual control enabled, use the instrument schematic to close gas inlet
valve.
10 Troubleshooting
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
10 - 3
Содержание GEMINI VII
Страница 1: ...OPERATOR MANUAL 239 42828 01 May 2021 Rev H GEMINI VII SURFACE AREA ANALYZER ...
Страница 10: ......
Страница 16: ......
Страница 92: ......
Страница 123: ...REPORT EXAMPLES BJH ADSORPTION REPORT Report Examples Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 31 ...
Страница 124: ...6 About Reports HORVATH KAWAZOE REPORT 6 32 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 125: ...ISOTHERM LINEAR PLOT REPORT Isotherm Linear Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 33 ...
Страница 126: ...6 About Reports SAMPLE LOG REPORT 6 34 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 127: ...T PLOT REPORT t Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 35 ...
Страница 128: ...6 About Reports VALIDATION REPORT 6 36 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 234: ......
Страница 242: ......
Страница 244: ......
Страница 247: ......
Страница 248: ......