Field or Button
Description
Include All.
Select to include all pressure points in the table.
Insert Predefined.
Click to insert a predefined (default) set of points
into the report.
Use Interpolation
must be selected to enable this
button. This button displays for BET reports only.
Use Interpolation.
Use to indicate if the system should use the table
or interpolated data. This option is available for BET and Langmuir
reports only.
Select Pressure Range
for BET fit [
text box
]
Enter values to indicate the fitted pressure range.
Selected Reports
[
group box
]
BET Isotherm plot.
Uses BET monolayer capacity and constant to
produce an isotherm.
n
Autoscale x-axis
. Linear x-axes begin at zero. The x-axis field
shows the relative pressure for BET.
n
Autoscale y-axis
. The y-axis field shows the quantity of gas
adsorbed.
n
Overlay samples.
Use to overlay sample files on the
BET isotherm plot.
BET Transform plot.
Use to generate a traditional BET surface
area plot used to determine monolayer volume and BET C constant.
n
Autoscale x-axis
. Linear x-axes begin at zero. The x-axis field
shows the relative pressure for BET.
n
Autoscale y-axis
. The y-axis field shows BET transformation.
n
Overlay samples.
Use to overlay sample files on the
BET transform plot.
Tabular report.
Use to have a table of measured and calculated
values generated.
For fields and buttons not listed in this table, see
.
BET Surface Area Report (continued)
BET Surface Area Report
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
7 - 5
Содержание GEMINI VII
Страница 1: ...OPERATOR MANUAL 239 42828 01 May 2021 Rev H GEMINI VII SURFACE AREA ANALYZER ...
Страница 10: ......
Страница 16: ......
Страница 92: ......
Страница 123: ...REPORT EXAMPLES BJH ADSORPTION REPORT Report Examples Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 31 ...
Страница 124: ...6 About Reports HORVATH KAWAZOE REPORT 6 32 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 125: ...ISOTHERM LINEAR PLOT REPORT Isotherm Linear Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 33 ...
Страница 126: ...6 About Reports SAMPLE LOG REPORT 6 34 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 127: ...T PLOT REPORT t Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 35 ...
Страница 128: ...6 About Reports VALIDATION REPORT 6 36 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Страница 234: ......
Страница 242: ......
Страница 244: ......
Страница 247: ......
Страница 248: ......