23
Technical specifications
Operation temperature:
•
10 to 40 °C (50 to 104 °F)
Storage temperature/
humidity:
•
-20 to 55 °C (-4 to 131 °F)
•
15 - 90 % relative maximum humidity
Weight:
•
625 g (including rechargeable battery
pack)
Dimensions:
•
110 x 177 x 175 mm
Measuring method:
•
Oscillometric, corresponding to
Korotkoff
Measurement range:
•
30 - 280 mmHg – blood pressure
•
40 - 200 beats per minute – pulse
Cuff pressure display:
•
Range: 0 - 299 mmHg
•
Resolution: 1 mmHg
•
Static accuracy: pressure within ± 3
mmHg
•
Pulse accuracy: ±5 % of the readout
value
Voltage source:
•
Rechargeable battery pack;
4.8V, 2400 mAh
•
Mains adapter DC 7.5V, 2 A
Reference to Standards:
• Device corresponds to the
requirements of the standard for
noninvasive blood pressure monitor.
EN 1060-1
EN 1060-3
EN 1060-4
IEC 60601-1
IEC 60601-1-2
Electromagnetic
compatibility:
• Device fulfills the stipulations of the
standard IEC 60601-1-2.
The stipulations of the EU Directive 93/42/
EEC for Medical Devices Class IIa have been
fulfilled.
Type BF applied part
Microlife reserves the right to alter technical
specfications without prior written notice.