Microlife BP W70 BT Скачать руководство пользователя страница 22

BP3KN1-3B 

 

22 

IB BP3KN1-3B V4920 EN   
   

EN

12. Technical specifications 

Operating conditions:

 ………10 - 40 °C / 50 - 104 °F 

15 - 90 % relative maximum humidity 

Storage conditions:

 …………-20 - +55 °C / -4 - +131 °F 

15 - 90 % relative maximum humidity 

Weight:

 …………………………120 +/-5 g (including batteries) 

Dimensions:

 ……………………82 x 64 x 13.9 mm 

Cuff size:

 ………………………13.5 – 23cm (5.25 – 9.06 inches) 

Measuring procedure:

 ………Oscillometric, corresponding to Korotkoff method: Phase I 

systolic, Phase V diastolic 

Measurement range:

 …………

SYS: 60~255 mmHg 

DIA: 40~200 mmHg   

40 - 199 beats per minute – pulse 

Cuff pressure display range:

 0 - 299 mmHg 

Resolution:

 ………………………1 mmHg 

Static accuracy:

 ……………… pressure within ± 3 mmHg 

Pulse accuracy:

 …………………± 5 % of the readout value 

Voltage source:

 …………………2 x 1.5 V alkaline batteries; size AAA 

Battery lifetime:

 …………………approx. 300 measurements (using new batteries) 

IP Class:

 ………………………… IP22 

Expected service life:

 ………… Device: 5 years or 10000 measurements   

Accessories: 2 years 

Atmospheric pressure range

…70KPa -106KPa 

Accessories included

………user manual

Storage Box

Gift box

BP diary

2xAAA 

Batteries 

 

Technical alterations reserved. 

13. How to contact us 

Distributed by: 

Microlife USA, Inc. 

1617 Gulf to Bay Blvd. 

Clearwater, FL 33755 

Toll Free Help Line: 1-800-568-4147 

Email: [email protected] 

www.microlifeusa.com 

Содержание BP W70 BT

Страница 1: ...BP3KN1 3B 1 IB BP3KN1 3B V4920 EN EN Microlife BP W70 BT BP3KN1 3B User Manual English...

Страница 2: ...ch from the hand This blood pressure monitor IS NOT intended for use with Pregnancy Preeclampsia Children under 12 Neonatal patients WARNING Potentially hazardous situation that if not avoided may res...

Страница 3: ...zardous situation that if not avoided may result in minor moderate injury property damage and or damage to the device This device contains sensitive electronics components Avoid strong electrical or e...

Страница 4: ...against vertically falling water drops when the device is tilted up to 15 Type BF Applied Part Batteries and electronic instruments must be disposed of in accordance with the locally applicable regula...

Страница 5: ...The device has been evaluated to meet general RF exposure requirement The device can be used in portable exposure condition without restriction Trademark Usage Apple the Apple logo iPad and iPhone are...

Страница 6: ...monitor 4 How do I get started 4 1 Inserting the batteries 4 2 Selecting standard or MAM mode 4 3 Selecting Heart Level Indicator 4 4 Setting the date and time 4 5 Setting the user 5 How do I take a...

Страница 7: ...ruction manual carefully to ensure correct use If you have additional questions regarding blood pressure measurements please contact your doctor 1 2 Inflation mode technology Please note that this dev...

Страница 8: ...above 100 mmHg while at rest In this case please consult your physician immediately Long term values at this level endanger your health due to continual damage to the blood vessels in your body If you...

Страница 9: ...n can be ascribed to specific organ malfunctions Please consult your physician for information about the possible origins of your own increased blood pressure values d There are measures you can take...

Страница 10: ...Battery Compartment 6 Cuff 7 Systolic Value 8 Diastolic Value 9 Pulse Rate 10 Date Time 11 Stored Value 12 Traffic Light Indicator 13 Irregular Heartbeat IHB Symbol 14 Cuff Signal Indicator ERR 1 15 C...

Страница 11: ...sure they are correctly connected as shown on the symbols in the compartment If the blood pressure monitor is left unused for long periods remove the batteries from the device 4 2 Selecting standard o...

Страница 12: ...LED is lighted with a heart symbol Conversely 1 or 2 upward arrow s appear when your wrist height is lower to heart level and raise the wrist higher as indication Adjust your wrist to correct height l...

Страница 13: ...that your wrist is free Always ensure that the cuff is positioned correctly as shown in the pictures illustrated on the short instruction card Fit the cuff comfortably but not too tight The cuff will...

Страница 14: ...the ON OFF button e g if you feel uneasy or an unpleasant pressure sensation 5 4 Stopping a measurement If it is necessary to interrupt a blood pressure measurement for any reason e g the patient feel...

Страница 15: ...optimal cuff fit 15 A are not considered in the average value Memory full Pay attention that the maximum memory capacity of 99 memories is not exceeded When the 99 memory is full the oldest value is a...

Страница 16: ...and time must be reset the year number therefore flashes automatically after the batteries are replaced after MAM mode and Heart Level Indicator setting Which batteries and which procedure Use 2 new...

Страница 17: ...minutes until a connection occurs To manually activate the Bluetooth on your monitor press and hold the power button for approximately 4 seconds or until the Bluetooth icon starts blinking If a conne...

Страница 18: ...ate and no result can therefore be displayed Read through the checklist for performing reliable measurements and then repeat the measurement ERR 6 MAM Mode There were too many errors during the measur...

Страница 19: ...ently fails to measure blood pressure values or the values measured are too low too high 1 Check the positioning of the cuff 2 Measure blood pressure again in peace and quiet carefully following the d...

Страница 20: ...p a minimum distance of 3 3 m from such devices when using this device Do not use this device if you think it is damaged or notice anything unusual Never open this device If the device is not going to...

Страница 21: ...improper handling and accidents Professional use not following the operating instructions and alterations made to the monitor or accessory by third parties are also not included in this warranty Some...

Страница 22: ...pulse Cuff pressure display range 0 299 mmHg Resolution 1 mmHg Static accuracy pressure within 3 mmHg Pulse accuracy 5 of the readout value Voltage source 2 x 1 5 V alkaline batteries size AAA Battery...

Отзывы: