9

5.7. Memory – displaying the last measurement

The blood-pressure monitor automatically stores the last
measurement value. By pressing the MEMORY button for
at least 3 seconds, the last measurement will be 
displayed.

Further information

Measurements should not occur soon after each other, since otherwise, the results will be falsified. Wait
therefore for at least 

1 minute

in a relaxed position, sitting or lying, before you repeat a measurement.

6.

Error messages/malfunctions

If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(example: error no. 1)

Error No.

Possible cause(s)

ERR 1

The systolic pressure was determined but afterwards the cuff pressure fell below
20 mmHg. This situation can occur, for example, if the tube becomes unplugged af-
ter the systolic blood-pressure has been measured.
Further possible cause: No pulse has been detected

ERR 2

The pressure dropped below 20 mmHg without a measurement taking place. The
Release valve was pressed during the measurement.

ERR 5

The measured readings indicated an unacceptable difference between systolic and 
diastolic pressures. Take another reading following directions carefully. Contact 
you doctor if you continue to get unusual readings. 

d)

After the inflation process lay down the pump ball
and remain in a quiet and relaxed sitting position. The
measurement now proceeds on its own. If no measu-
rement takes place and all that appears in the display
is a flashing, upwards pointing arrow, see below 
under «Inflate further». When the instrument detects
a pulse, a heart symbol in the display starts to flash
and a beep is heard for every heart beat.

e)

When the measurement has been concluded, a long
peep tone sounds. The measured systolic and 
diastolic blood-pressure values as well as the pulse
frequency are now displayed.
The cuff pressure can now be quickly released by
pressing the release valve at the pump ball.

5.5. Inflating further

If the cuff is not inflated sufficiently, the measurement is
broken off after a few seconds and a flashing arrow 
pointing upwards appears in the display.
It is then necessary to inflate further to at least 20 mmHg
higher than the previous value.
The instruction to inflate further can appear several times
if the pressure is still not sufficient.

Attention:

If the cuff is inflated too far (over 300 mmHg), «HI» 
appears in the display and a warning signal is sounded.
The cuff pressure must then be immediately released
using the quick-release valve) at the pump ball)! 
As soon as the cuff is inflated sufficiently and the 
measurement is proceeding, continue as described under
points d–e.

5.6. Discontinuing a measurement

If it is necessary to interrupt a blood pressure measure-
ment for any reason (e.g. the patient feels unwell), press
the quick release valve at the pump ball. The device then
immediately lowers the cuff-pressure automatically.

8

3 sec.

Содержание BP 3BTO-H

Страница 1: ...fe BP 3BTO H Semi automatic Blood Pressure Monitor Instruction Manual 1 13 14 29 P automatyczny ci nieniomierz krwi Instrukcja obs ugi 30 43 F lautomata v rnyom sm r Haszn lati tmutat 44 57 www microl...

Страница 2: ...Czech Republic Tel 420 56 7212 396 Fax 420 56 7212 396 Estonia Oriola AS Estonia Saku tn 8 EE 11314 Tallinn Estonia Tel 372 65 15 100 Fax 372 65 15 111 Greece Cyprus Karabinis ALFA GAUZE Ltd 151 Lavr...

Страница 3: ...deposits in the vessel walls Arteriosclerosis A deficient supply of blood to important organs heart brain muscles can be the result Furthermore with long term continuously increased blood pressure va...

Страница 4: ...n the common primary essential hypertension and secondary hypertension The latter group can be ascribed to specific organic malfunctions Please consult your doctor for information about the possible o...

Страница 5: ...sely to your upper arm Measure always on the same arm normally left Attempt to carry out the measurements regularly at the same time of day since the blood pressure changes during the course of the da...

Страница 6: ...ng position The measurement now proceeds on its own If no measu rement takes place and all that appears in the display is a flashing upwards pointing arrow see below under Inflate further When the ins...

Страница 7: ...and then exchange them for new The pressure does not rise although the Check the connection of the cuff tube and pump is running connect properly if necessary The device frequently fails to measure t...

Страница 8: ...ircumference 22 32 cm cuff type L for arm circumference 32 42 cm Technical alterations reserved 11 www microlife com Detailed user information about our thermometers and blood pressure monitors as wel...

Страница 9: ...15 1 1 1 1 BP 3 1 2 2 2 2 1 K 2 2 K 2 3 3 3 C C 4 4 4 1 4 2 5 5 5 1 5 2 5 3 5 4 5 5 5 6 5 7 6 6 7 7 8 8 9 9 1 10 0 1 11 1 w ww ww w m miic cr ro olliif fe e c co om m...

Страница 10: ...17 1 1 1 1 1 1 B BP P 3 3 x 1 1 2 2 2 2 2 2 1 1 K K 16 2 2 2 2 K K K 90 160 140 160 90 100 K K 110 60 mmHg 3 100 60 100 140 60 90 140 160 90 100 160 180 100 110 180 110...

Страница 11: ...K K 40 3 3 BP 3 K K T M 22 32 L 32 42 19 120 2 2 3 3 K A A Diabetes mellitus 18 AC 1...

Страница 12: ...21 4 4 4 4 1 1 AA 4 316 R6 1 5 1 5 1 2 K K I O 4 4 2 2 1 1 1 20 5 5 5 5 1 1 10 5 5 2 2 K K K 15 10 1 5 5 3 3...

Страница 13: ...23 2 3 2 5 5 4 4 I 2 t 0 22 40 160 180 5 5 5 5 5 5 s 20 300 HI 2 3 cm letku...

Страница 14: ...25 5 5 6 6 5 5 7 7 MEMORY 1 6 6 1 E ER RR R 1 1 20 E ER RR R 2 2 E ER RR R 5 5 15 RWT V Essen 3 sec 24 1 2 1 2 3...

Страница 15: ...27 7 7 26 8 8 BP 3BTO H 9 9 EN1060 1 12 95 EN1060 3 09 97 DIN 58130 NIBP clinical investigation ANSI AAMI SP10 NIBP requirements EN 60601 1 2 K K DIN 58130 1997 6 93 42 EWG IIa...

Страница 16: ...29 1 10 0 265 131 174 73 5 50 C 15 85 10 40 C 30 280 40 200 0 299 1 3 5 40 200 4 316 R6 1 5 K K 22 32 L 32 42 1 11 1 w ww ww w m mi ic cr ro ol li if fe e c co om m www microlife com 28...

Страница 17: ...ci nienia 3 Elementy ci nieniomierza 4 Obs uga ci nieniomierza 4 1 Zak adanie baterii 4 2 Pod czanie w a gumowego 5 Przeprowadzenie pomiaru 5 1 Przed pomiarem 5 2 Najcz stsze r d a b d w 5 3 Zak adani...

Страница 18: ...wnym chorobom warto ci ci nienia musz si mie ci w okre lonych granicach normy 32 33 2 2 Jakie warto ci ci nienia s normalne O zbyt wysokim ci nieniu m wimy w wczas gdy mierzone w spokoju ci nienie roz...

Страница 19: ...arza dotycz cymi przebytych chor b takich jak cukrzyca zaburzenia metabolizmu t uszcz w dna moczanowa Na ogi Rzucenie palenia U ywanie tylko umiarkowanych ilo ci alkoholu Ograniczenie spo ycia kofeiny...

Страница 20: ...arze o 10 mmHg 36 37 Mankiety kt re s zbyt w skie lub zbyt kr tkie prowadz do uzyskania niewiarygodnych wynik w Dob r w a ciwego mankietu jest wi c bardzo wa ny Rozmiar mankietu zale y od obwodu ramie...

Страница 21: ...jawia si na ekranie symbol serca i s ycha pisz cz cy sygna pojawiaj cy si przy ka dym uderzeniu serca 38 39 e Po zako czeniu pomiaru s ycha d ugi sygna Mo na teraz wypu ci powietrze z mankietu przez n...

Страница 22: ...Dalsze wskaz wki Poziom ci nienia krwi podlega wahaniom nawet u zdrowych os b Dlatego nale y pami ta e por wnywane pomiary wymagaj zawsze takich samych warunk w spokoju Je li jednak przy zastosowaniu...

Страница 23: ...on przez dystrybutora Nazwa i adres dystrybutora 42 43 9 Standardy Standard aparatu Aparat odpowiada wymogom standardu europejskiego dla ci nieniomierzy nieinwazyjnych EN1060 1 12 95 EN1060 3 09 97 DI...

Страница 24: ...a lland an magas vagy alacsony a v rnyom sa 3 A v rnyom sm r k sz l k k l nb z r szei 4 A v rnyom sm r k sz l k zembe helyez se 4 1 Az elemek behelyez se 4 2 A mandzsetta vezet k csatlakoztat sa 5 A m...

Страница 25: ...onzul t ljon orvos val Ha a v rnyom sa huzamos ideig ilyen magas ez vesz lyeztetheti az eg szs g t mivel a testben l v v red nyek k rosod s hoz vezethet 46 Ha a szisztol s rt k 140 s 160 Hgmm k z tt v...

Страница 26: ...z al bbi bra a BP 3BTO H t pus v rnyom sm r k sz l k r szeit mutatja be a M r eszk z 48 b Mandzsetta opci M m ret 22 32 cm es karker letre L m ret 32 42 cm es karker letre k l nleges tartoz kk nt megv...

Страница 27: ...al bb h rom percig felemelve kell tartani hogy elfolyj on az sszegy lt v r 5 3 A mandzsetta el k sz t se a A mandzsetta v g t vezesse t a levarrt gumi v g vel a f m kengyelen kereszt l gy hogy hurkot...

Страница 28: ...r s le ll t sa Ha b rmilyen okn l fogva pl ha a beteg rosszul rzi mag t nyomja meg a pump n tal lhat leereszt sze lep gombj t Ekkor a nyom s azonnal cs kkeni kezd a mandzsett ban 52 5 7 Mem ria funkc...

Страница 29: ...egyes k sz l kben s ezeket ugyan gy klinikailag tesztelj k A k sz l k gy rt ja elfogadja s a k sz l k maga megfelel a v rnyom sm r k sz l kek eur pai szabv ny gyi el r sainak lsd a technikai adatokat...

Страница 30: ...131 x 174 x 73 mm T rol si h m rs klet 5 s 50 C k z tt P ratartalom 15 85 relat v p ratartalom zemeletet si h m rs klet 10 s 40 C k z tt Kijelz Folyad kkrist lyos LCD M r si m d Oszcillometrikus Nyom...

Отзывы: