IM-LIPOTOWER Rev G

7

Rating: 100-120V ~ 60 Hz 1200VA 

Maximum suction: 27-29 inHg (685.8-736.6 mmHg) at sea level

Output: 360VA   

 

 

Maximum continuous run time: 90 minutes

SETUP

WARNING: 

To avoid personal injury, always 

ensure that the system is disconnected from 
applicable energy sources before starting any 
procedure in this section.

CAUTION: 

Some components may be 

Electrostatic Sensitive.  Observe precautions for 
handling Electrostatic Sensitive Devices.

1. LOCKING AND UNLOCKING CASTERS

1.  The front casters can be locked by 

depressing the brake tab that is 
connected to the caster. 

(Figure 1)

2.  The front casters can be unlocked by 

lifting the brake tab that is connected 
to the caster. 

(Figure 2)

NOTE:  

Casters should be locked during set up 

and operation.

2. CONNECTING A MICROAIRE POWER 
CONSOLE FOR PAL SYSTEM (OPTIONAL)

1.  Place Electric Console (REF 1020/1025 

or REF 5020) on the shelf. 

(Figure 3)

2.  Remove the back cover panel on the 

upper rear of the cart by loosening the 
two retaining screws. 

(Figures 4 and 5)

3.  Manuever the electrical cord from the REF 1020/1025 or REF 5020 through 

the cable opening and connect it to the hospital grade outlet on the back 
of the cart. 

(Figure 6)

4.  Manuever the BNC cable from the REF 1020/1025 or REF 5020 through the 

cable opening and connect it to the port on the back of the cart. 

(Figure 7)

5.  Replace the rear cover and tighten the two retaining screws.

CAUTION: 

The Hospital Grade Receptacle has maximum current limit of 3A.   

Do not exceed.

CAUTION: 

This panel is in place for safety. Only UL-approved, PAL-related MicroAire cables are to be used to connect to 

the receptacles within the system.  Ensure cover is securely attached before connecting power.

TECHNICAL DESCRIPTION

The MicroAire LipoTower System is Class I, Type B equipment, designed for continuous operation. It should not be 
used in the presence of flammable anesthetics. This equipment is rated IPX0 and provides no protection against the 
ingress of solids or liquids.    

This equipment provides suction for health care facilities and is not intended for use in the field or during transport.

Figure 1 - Lock the casters

Figure 3 - Install optional console 

Figure 4 - Unscrew back-cover screws

Figure 5 - Remove back panel

Figure 6 - Plug console into outlet

Figure 7 - Attach BNC connector

Figure 2 - Unlock the casters

Содержание ASP-1021

Страница 1: ...MicroAire LipoTower System Instructions for Use...

Страница 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Страница 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Страница 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Страница 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Страница 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Страница 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Страница 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Страница 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Страница 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Страница 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Страница 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Страница 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Страница 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Страница 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...

Страница 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...

Страница 17: ...IM LIPOTOWER Rev G 17 NOTES...

Страница 18: ...IM LIPOTOWER Rev G 18 NOTES...

Страница 19: ...IM LIPOTOWER Rev G 19 NOTES...

Страница 20: ...nstruments LLC IM LIPOTOWER Rev G Printed in the USA M MicroAire Surgical Instruments LLC 3590 Grand Forks Blvd Charlottesville VA 22911 USA Telephone 800 722 0822 434 975 8000 Fax 800 648 4309 434 97...

Отзывы: