IM-LIPOTOWER Rev G
12
INSTALLING AN EXTRA SHELF
1. Turn off power and disconnect main power by unplugging power cord.
2. Insert the shelf from the front of the cart toward the rear and line up the bolt holes.
3. Insert the four bolts and use a 5/32” Allen wrench to tighten the bolts until they are snug.
TROUBLESHOOTING
DIFFICULTY INSERTING CABLES
•
Align connectors and receptacles carefully.
•
Make sure that all markings on cable plugs are aligned with matching markings on receptacles.
•
If difficulty persists, contact MicroAire Customer Service for additional troubleshooting and repair.
THE MICROAIRE LIPOTOWER SYSTEM WILL NOT POWER UP
•
Position the MAINS DISCONNECT SWITCH, located on the rear of the MicroAire LipoTower System to the OFF
position. Ensure that the power cord is fully seated in the wall outlet. Position the MAINS DISCONNECT SWITCH
to the ON position.
•
Verify that the STANDBY switch is in the ON position.
•
If the unit is still not operating properly, contact MicroAire Customer Service for additional troubleshooting
and repair.
FOOT SWITCH WILL NOT OPERATE
•
Ensure that the foot switch connection tube is fully seated in its receptacle at the lower rear of the MicroAire
LipoTower System.
•
Ensure that the vacuum pump CONTROL SWITCH is in the FOOT SWITCH position.
•
If the foot switch still does not operate properly, contact MicroAire Customer Service for additional
troubleshooting and repair.
VACUUM PRESSURE IS INSUFFICIENT
•
Ensure all aspiration canisters and aspiration lines are properly connected and not kinked, and all aspiration
canister lids are properly installed, with no clogged filters.
•
Check the biofilter to ensure it is not clogged.
•
Ensure that all applicable component connections are secure (the foot switch, REF PAL-650 handpiece, REF
1020/1025 and REF 5020 electric console); and that the 3-WAY CONTROL SWITCH is properly set.
•
If vacuum pressure is still insufficient, contact MicroAire Customer Service for additional troubleshooting and
repair.
MICROAIRE LIPOTOWER SYSTEM CIRCUIT BREAKER IS TRIPPED
•
Check the biofilter to ensure it is not clogged.
•
Cycle the MAINS DISCONNECT SWITCH to the OFF position and then to the ON position.
•
If the Circuit Breaker trips again, contact MicroAire Customer Service for additional troubleshooting and repair.
CLEANING AND STERILIZATION
WARNING:
The MicroAire LipoTower System should never be sterilized, immersed, or washed.
Before cleaning, the MicroAire LipoTower System should be unplugged. External surfaces of the device should be
carefully wiped down with a disinfectant after each procedure, and at the beginning of each day. Care should be taken
not to drip fluid into any equipment openings.
WARNING:
If any fluid or solid material is drawn into the vacuum pump, the equipment must be serviced.
Содержание ASP-1021
Страница 1: ...MicroAire LipoTower System Instructions for Use...
Страница 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...
Страница 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...
Страница 17: ...IM LIPOTOWER Rev G 17 NOTES...
Страница 18: ...IM LIPOTOWER Rev G 18 NOTES...
Страница 19: ...IM LIPOTOWER Rev G 19 NOTES...