IM-LIPOTOWER Rev G

4

WARNING:

 Explosion hazard. Do not use in the presence of flammable anesthetics or oxygen.

WARNING: 

The MicroAire LipoTower System is heavy, weighing approximately 190 pounds.  Use the proper equipment 

to lift the system if required. 

WARNING: 

Flat shelving capacity – 6.8 kg (15lb) maximum safe capacity.  Do not exceed.

WARNING:

  Tumescent pole capacity - 3kg (6.6lb) maximum safe capacity per hook. Do not exceed.

WARNING:

  Do not transport cart with anything on shelves or hooks.  Remove all items from cart before  

unlocking casters.

WARNING: 

Do not modify any accessory. Failure to comply may result in patient and or operating room staff injury 

and equipment damage.

WARNING: 

Use only MicroAire-approved accessories. Use of other unapproved accessories may result in increased 

emissions or decreased immunity of the system and will void your warranty.  

WARNING: 

This device will not, in and of itself, produce significant weight reduction.

WARNING: 

This device should be used with extreme caution in patients with chronic medical conditions, such as 

diabetes, heart, lung or circulatory system disease or obesity.

WARNING: 

The volume of blood loss and endogenous body fluid loss may adversely affect intra and/or postoperative 

hemodynamic stability and patient safety.  The capability of providing adequate, timely replacement is essential for 

patient safety.

WARNING: 

Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle 

marked “Hospital Only” or “Hospital Grade.”

WARNING:

 Disconnecting the supply cord will isolate the system from the supply mains on all poles simultaneously.

WARNING:

 Inspect tubing and suction canisters for wear or damage before use.

WARNING: 

Risk of fire. Use only MicroAire cables to connect to this device.

WARNING: 

Electric shock. Do not remove cover; refer servicing to qualified personnel only.

WARNING: 

Use care to ensure that there is no electromagnetic interference between this device and other  

devices in use.

WARNING:  

Casters must be locked to prevent movement during normal use when a canister containing fluid is 

installed on the canister rack. Do not roll or tilt the cart in this condition because fluid could spill onto the cart causing 

a hazardous situation. 

WARNING: 

ESD (Electromagnetic environment – guidance): Floors should be wood, concrete or ceramic tile. If floors 

are covered with synthetic material, the relative humidity should be at least 30%. If ESD-related issues are experienced 

during use, it may be necessary to use antistatic materials (e.g., anti-static mat) or to increase the relative humidity of 

the surrounding environment.

Содержание ASP-1021

Страница 1: ...MicroAire LipoTower System Instructions for Use...

Страница 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Страница 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Страница 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Страница 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Страница 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Страница 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Страница 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Страница 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Страница 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Страница 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Страница 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Страница 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Страница 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Страница 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...

Страница 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...

Страница 17: ...IM LIPOTOWER Rev G 17 NOTES...

Страница 18: ...IM LIPOTOWER Rev G 18 NOTES...

Страница 19: ...IM LIPOTOWER Rev G 19 NOTES...

Страница 20: ...nstruments LLC IM LIPOTOWER Rev G Printed in the USA M MicroAire Surgical Instruments LLC 3590 Grand Forks Blvd Charlottesville VA 22911 USA Telephone 800 722 0822 434 975 8000 Fax 800 648 4309 434 97...

Отзывы: