MicroAire ASP-1021 Скачать руководство пользователя страница 5

IM-LIPOTOWER Rev G

5

WARNING:  

EFT (Electromagnetic environment – guidance): Mains power quality should be that of a typical 

commercial or hospital environment. If EFT-related issues are experienced during use, the use of power-line filtering 
may be necessary to ensure proper operation in the presence of large power-line transients.

WARNING:

 Do not block cooling vents. Maintain 50mm MIN clearance.

CAUTION: 

This device is designed to contour the body by removing localized deposits of excess fat through small 

incisions and subsequently transfer the tissue back to the patient.

CAUTION:

 Do not leave cannulas in patient when not in use.

CAUTION: 

Use of this device is limited to those physicians who, by means of formal professional training or sanctioned 

continuing medical education (including supervised operative experience), have attained proficiency in suction 
lipoplasty and tissue transfer.

CAUTION:

 Results of this procedure will vary depending on patient age, surgical site and experience of the physician.

CAUTION: 

Results of this procedure may or may not be permanent.

CAUTION:

 The amount of fat removed should be limited to that necessary to achieve a desired cosmetic effect.

CAUTION:

 All reusable components of the device must be sterilized and all disposable components replaced before 

using the device system on another patient.

CAUTION:

 Federal Law restricts this device to sale by, or on the order of, a physician (or properly licensed practitioner).

CAUTION:

 See “PAL-650 Instrument System Instruction Manual” for detailed information on the PAL-650 handpiece.

CAUTION:

 See “Model 1020 Standard Electric Console Instruction Manual” for detailed information on the Model 1020 

Standard Electric Console. See “Model 5020 Standard Electric Console Instruction Manual” for detailed information on 
the Model 5020 Standard Electric Console.

CAUTION: 

Do not use this device to aspirate surgical smoke from devices such as cauterizers, lasers, electrosurgical 

instruments or ultrasonic units.  Doing so may clog the biofilter.

NOTE: 

This device is intended only for the aspiration of tissue or fluid from the body.  The tissue product collected by 

this device shall only be used with FDA exempt/cleared/approved devices for any further use other than disposal.

NOTE: 

All personnel involved with the operation, cleaning, maintenance and cleaning of the REF ASP-1020, -1021, 

should become familiar with the power equipment before it is set up for use in any procedure. Personnel that are 
trained should include, but not be limited to, central processing personnel, members of the surgical team, and the 
bioengineering department.

NOTE: 

The MicroAire LipoTower System should only be used with compatible FDA cleared tubing, handpieces and 

cannulas.

NOTE: 

The MicroAire LipoTower System has been tested for sound level and does not exceed 70 dB(A).

Содержание ASP-1021

Страница 1: ...MicroAire LipoTower System Instructions for Use...

Страница 2: ...nsole for PAL REF 5025 Electric Power Console for MicroAire Instruments REF PAL 1200 Liposuction Tubing 12 foot REF 1006 PALE PAL Instrument Cable for REF 1020 1025 REF 5006 PAL PAL Instrument Cable f...

Страница 3: ...nd the 5020 5025 Electric Console are described in their respective IFUs This IFU addresses primarily the operation maintenance cleaning and sterilization procedures applicable to the MicroAire LipoTo...

Страница 4: ...nd or postoperative hemodynamic stability and patient safety The capability of providing adequate timely replacement is essential for patient safety WARNING Grounding reliability can only be achieved...

Страница 5: ...system on another patient CAUTION Federal Law restricts this device to sale by or on the order of a physician or properly licensed practitioner CAUTION See PAL 650 Instrument System Instruction Manual...

Страница 6: ...CH ON allows pumps to be activated by the Control Switch STANDBY o turns vacuum pumps off 6 STORAGE BASKET 7 LOCKING CASTERS 8 PUSH HANDLE 9 MAIN VACUUM CONNECTION 10 REAR STORAGE BASKET 11 POWER INLE...

Страница 7: ...20 1025 or REF 5020 through the cable opening and connect it to the hospital grade outlet on the back of the cart Figure 6 4 Manuever the BNC cable from the REF 1020 1025 or REF 5020 through the cable...

Страница 8: ...er Rack on the top of the cart Slide one or more Waste Canister Rings REF ASP RNG 1200 or REF ASP RNG 2000 into any of the four available canister slots These slots also accept various wall mount styl...

Страница 9: ...or to use all system components should be inspected to detect any damage excessive wear corrosion or malfunction DO NOT use if any potential problems become apparent Contact MicroAire for resolution W...

Страница 10: ...PRESSURE ADJUSTMENT knob controls the amount of suction pressure Vacuum pressure is displayed on the LCD screen Suction is increased by turning the knob clockwise towards MAX and suction is decreased...

Страница 11: ...N OPTIONAL INSTALLING THE TUMESCENT MODULE 1 Turn off power and disconnect main power by unplugging power cord 2 Locate the Tumescent Bay cover on the left side of the cart Remove the the two retainin...

Страница 12: ...the foot switch still does not operate properly contact MicroAire Customer Service for additional troubleshooting and repair VACUUM PRESSURE IS INSUFFICIENT Ensure all aspiration canisters and aspira...

Страница 13: ...roAire LipoTower System is operated at high elevation or low atmospheric pressure aspiration performance may be adversely affected NOISE LEVEL EXPOSURE LIMITS The A weighted emission sound pressure le...

Страница 14: ...SUMMARY CAUTION The MicroAire LipoTower system needs to be installed and put into service according to the EMC information provided in this Instructions for Use manual Portable and mobile RF communica...

Страница 15: ...IM LIPOTOWER Rev G 15 Table 1 IEC 60601 1 2 Table 201 Table 2 IEC 60601 1 2 Table 202...

Страница 16: ...IM LIPOTOWER Rev G 16 Table 4 IEC 60601 1 2 Table 206 Table 3 IEC 60601 1 2 Table 204...

Страница 17: ...IM LIPOTOWER Rev G 17 NOTES...

Страница 18: ...IM LIPOTOWER Rev G 18 NOTES...

Страница 19: ...IM LIPOTOWER Rev G 19 NOTES...

Страница 20: ...nstruments LLC IM LIPOTOWER Rev G Printed in the USA M MicroAire Surgical Instruments LLC 3590 Grand Forks Blvd Charlottesville VA 22911 USA Telephone 800 722 0822 434 975 8000 Fax 800 648 4309 434 97...